Day Surgery General Pack, FHGE18T recallZ-3231-2024 · American Contract Systems, Inc. · initiated Aug 7, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
FDA root cause category: Under Investigation by firm
What the company is doing
Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " -Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Discard all affected kits and document on the enclosed Response Form."
Recall details
- Recall number
- Z-3231-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 7, 2024
- Posted by the FDA
- Sep 24, 2024
- Recalling firm
- American Contract Systems, Inc.
- Quantity
- 280 units
- Product code
- LRO
- Distribution
- US Nationwide distribution in the states of MO, MN, MA, OH, NE.
Product description
Day Surgery General Pack, FHGE18T; Medical convenience kit
Code information (models, lots, serial numbers)
UDI-DI: 00191072210517; Lot: 2405131; Exp: 05/13/2025
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ACS FHGE18TAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95261).
- Z-3226-2024 · Open, ClassifiedPACEMAKER PACK, SLCV01J; Medical convenience kit
- Z-3227-2024 · Open, ClassifiedEndovascular AAA Pack, ESED52B; Medical convenience kit
- Z-3228-2024 · Open, ClassifiedEndovascular AAA Pack, ESED52B; Medical convenience kit
- Z-3229-2024 · Open, ClassifiedArthroscopy Pack, MSKA45A; Medical convenience kit
- Z-3230-2024 · Open, ClassifiedOpen Shoulder, LWOS39L; Medical convenience kit
- Z-3232-2024 · Open, ClassifiedShoulder, SESH18I; Medical convenience kit
- Z-3233-2024 · Open, ClassifiedBilateral Tubal Ligation Pack, FHBT3T; Medical convenience kit
- Z-3234-2024 · Open, ClassifiedParacentesis Thoracente, SLPT76H; Medical convenience kit
- Z-3235-2024 · Open, ClassifiedIR Procedure Pack, CEAT19W; Medical convenience kit
- Z-3236-2024 · Open, ClassifiedNeuro Vascular, ANKV91A; Medical convenience kit
- Z-3237-2024 · Open, ClassifiedHand Pack, AKHD97C; Medical convenience kit
- Z-3238-2024 · Open, ClassifiedVaginal Delivery, ASDV25B; Medical convenience kit
- Z-3239-2024 · Open, ClassifiedCath Lab Pack, CECL02P; Medical convenience kit
- Z-3240-2024 · Open, ClassifiedEssentia MAJ Laparotomy Std Pack, ESLT53B; Medical convenience kit
- Z-3242-2024 · Open, ClassifiedNasal Sinus Pack, MTSN26B; Medical convenience kit
- Z-3243-2024 · Open, ClassifiedCysto, FTCY03L; Medical convenience kit
- Z-3244-2024 · Open, ClassifiedPacemaker, COPM11B; Medical convenience kit
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page