Salem Sump recallZ-3302-2024 · Cardinal Health 200, LLC · initiated Jul 24, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
FDA root cause category: Labeling Change Control
What the company is doing
The recalling firm issued letters dated 7/22/2024 via overnight mail on 7/24/2024. The letter informed the consignee the product labeling has been updated via electronic instructions for use for the affected product due to reported incidences of improper use of the device which can lead to breakage of the Salem Sump Anti-Reflux Valve (ARV) and an increased risk to patients. To mitigate the risk of improper use, the instructions have been revised. The risk to health and actions required were included in the letter. The actions included: (1) Review your inventory for the affected product codes; (2) Communicate the change of the use instructions with all personnel that utilize the devices; (3) Post a copy of Attachment A (the revised IFU) and the recall notification in the storeroom and clinical areas; (4) Notify any customers to whom you may have distributed/forwarded affected product or to whom you intend to distribute/forward product) regarding this medical device product correction and share a copy of the notice; (5) Return the enclosed acknowledgment form via FAX or email.
Recall details
- Recall number
- Z-3302-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 24, 2024
- Posted by the FDA
- Sep 26, 2024
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- 24,761,100 tubes total
- Product code
- FEG
- FDA submissions
- K880850
- Distribution
- Distribution was nationwide, including Puerto Rico and Guam. There was also government/military distribution. Foreign distribution was made to Chile.
Product description
(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888266122, Sterile; (c) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) x 48 in. (122 cm), Model #8888266130, Sterile; and (d) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 18 Fr/Ch (6.0 mm) x 48 in. (122 cm), Model #8888266148, Sterile.
Code information (models, lots, serial numbers)
All lot numbers. UDI-DI numbers: (a) 8888266114, UDI-DI 10192253012781; (b) 8888266122, UDI-DI 10192253012804; (c) 8888266130, UDI-DI 10192253012828; and (d) 8888266148, UDI-DI 10192253012842.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Salem Sump 8888266114Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION2 recalls0 reports
- Salem Sump 8888266148Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall5 reports
- Salem Sump 8888266122Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall2 reports
- Salem Sump 8888266130Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95214).
- Z-3301-2024 · Open, Classified(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888264929, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube, 14 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page