Skip to content

GE Heathcare Giraffe OmniBed recallZ-0047-2025 · Datex--Ohmeda, Inc. · initiated Sep 3, 2024

Class I Open, Classified All recalls posted in October 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datex--Ohmeda, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

FDA root cause category: Use error

What the company is doing

GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 09/03/2024 via Fed-Ex. The notice explained the problem with the device, safety issue, and requested the following: Download the Service Manual Addendum (5971733) from the GE HealthCare Customer Documentation Portal: https://www.gehealthcare.com/support/manuals. Inspect the heater door mounting screw on your device. If the screw is not appropriately tightened, complete steps 9 and 10 of Section 1.1 of the Service Manual Addendum (5971733).

Recall details

Recall number
Z-0047-2025
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Sep 3, 2024
Posted by the FDA
Oct 17, 2024
Recalling firm
Datex--Ohmeda, Inc.
Product code
FMZ
FDA submissions
K101788, K020543
Distribution
Worldwide distribution.

Product description

GE Heathcare Giraffe OmniBed; neo natal incubator

Code information (models, lots, serial numbers)

All models

Other devices under the same FDA submission

1 device cleared under K101788, K020543 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95282).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page