IMMAGE recallZ-0118-2025 · Beckman Coulter Inc. · initiated Jul 31, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm identified two lots of reagents that are not meeting the sensitivity statement in the IFU
FDA root cause category: Employee error
What the company is doing
During the week of 08/15/2024, the firm an "URGENT MEDICAL DEVICE RECALL" Letter via first class mail and/or email. The letter informed customer that Beckman Coulter discovered that IMMAGE Immunochemistry Systems IGM Immunoglobulin M reagent lots are not meeting the sensitivity statement in the IFU: Sensitivity is defined as the lowest measurable concentration which can be distinguished from zero with 95% confidence. Sensitivity for IGM determination is 4.2 mg/dL. . Customers are instructed to: -Discontinue use of affected IMMAGE Immunochemistry Systems IGM Immunoglobulin M reagent lots and to follow product replacement instructions. -Review the Safety Data Sheet (SDS) and discard according to their laboratory's Standard Operating Procedures and/or their local regulations. SDS is available on www.beckmancoulter.com -Share the content of Recall Letter with their lab and/or medical director regarding the need to perform a retrospective review of patient results. Beckman Coulter is no longer distributing the affected lots. For questions, contact Customer Support Center: . From website: http://www.beckmancoulter.com . By phone: call 1-800-854-3633 in the United States. o Outside the United States, contact local Beckman Coulter representative. . For customers in the United States, if need replacement product: o Complete the attached Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821 . For customers in other geographies, contact local Beckman Coulter Representative for replacement.
Recall details
- Recall number
- Z-0118-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2024
- Posted by the FDA
- Oct 17, 2024
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 12,102 reagents
- Product code
- DEW
- FDA submissions
- K963868
- Distribution
- U.S.: AZ, CA, CO, FL, GA, HI, IL, MA, MI, MO, ND, NY, OH, PA, and TX O.U.S.: Australia, Belgium, Brazil, Chile, China, Czechia, Ecuador, France, Germany, Greece, Hong Kong, India, Italy, Kuwait, Mexico, Philippines, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan Province of China, and Turkey,
Product description
IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent REF 447610 IGM reagent, when used in conjunction with IMMAGE Immunochemistry Systems and Calibrator 1, is intended for the quantitative determination of immunoglobulin M (IGM) in human serum by rate nephelometry.
Code information (models, lots, serial numbers)
Catalog Number: 447610 UDI-DI code: 15099590221195 Lot Numbers: M307362 M309410
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IMMAGE 447610Beckman Coulter, Inc. · Method, Nephelometric, Immunoglobulins (G, A, M)1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page