IntelliVue Patient Monitor MX400 (Model Number 866060)… recallZ-3249-2024 · Philips North America · initiated Aug 30, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An URGENT Medical Device Correction notification letter dated 8/30/24 was sent to customers. Actions that should be taken by the customer / user to prevent risks for patients or users " Please check if the equipotential connector is broken off (refer to Figure 1). If yes, remove the device from use. " Pass this notice to all those who need to be aware within your organization or to any organization where affected devices have been potentially transferred. " Customers should complete the URGENT Medical Device Response Form either via the QR code below or at the end of the notification to submit both their acknowledgment of this recall and confirm understanding of actions to be taken. If hard copy of Response Form is completed in lieu of digital version (via QR code) please return to Philips at recall.responsephilips.com. You may also fax it to 877 499 7223. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will replace the power supply. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377. Distributors: Philips is requesting customers to return a Response Form to acknowledge receipt and understanding of the URGENT Medical Device Correction and confirm that the information from this Letter has been properly distributed to all users that handle the affected product. Together with this letter we are providing a list of affected products that Philips has sold to your organization. As distributor of the affected products, we kindly request that you: o Add in the Response Form attached your contact information. o Send the attached URGENT Medical Device Correction to each customer to whom you have distributed any affected product as soon as possible and no later than three days, together with the Reply Card. o Perform a good faith effort to get the Reply Form by following up with the customer with a minimum of t
Recall details
- Recall number
- Z-3249-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 30, 2024
- Posted by the FDA
- Sep 25, 2024
- Recalling firm
- Philips North America
- Quantity
- 545 devices
- Product code
- MHX
- FDA submissions
- K130849, K131872
- Distribution
- US Nationwide. Global Distribution.
Product description
IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
Code information (models, lots, serial numbers)
IntelliVue Patient Monitor MX400, 866060, UDI-DI: 00884838038752; IntelliVue Patient Monitor MX430, 866061, UDI-DI: 00884838057562; IntelliVue Patient Monitor MX450, 866062, UDI-DI: 00884838038769; IntelliVue Patient Monitor MX500, 866064, UDI-DI: 00884838038776; IntelliVue Patient Monitor MX550, 866066, UDI-DI: 00884838038783; Only MX400-550 devices shipped after 26-April-2024 are affected. Please refer to the manufacturing date on the back of your monitor.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- IntelliVue MX450 Patient Monitor IntelliVue Patient Monitor MX450Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)5 recalls22 deaths320 reports
- IntelliVue MX550 Patient Monitor IntelliVue Patient Monitor MX550Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)5 recalls12 deaths256 reports
- IntelliVue MX500 Patient Monitor IntelliVue Patient Monitor MX500Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)5 recalls6 deaths133 reports
- IntelliVue MX400 Patient Monitor IntelliVue Patient Monitor MX400Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)4 recalls6 deaths93 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page