IntelliVue G7m Anesthesia Gas Module, Product Number 866173 recallZ-0007-2025 · Philips North America · initiated Jan 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
FDA root cause category: Under Investigation by firm
What the company is doing
An IMPORTANT PRODUCT NOTICE dated 1/18/23 was sent to customers. The actions planned by Philips to correct the problem A Philips representative will reach out to you to arrange a replacement of the sample gas pump. Affected products and how to identify them Affected product is the 866173 lntelliVue G7m Anesthesia Gas Module. Both the product number and the serial number on the G7m product is contained on the back of the product, preceded by "REF" and "SN", respectively. If you need any further information or support concerning this issue, please contact your local Philips representative.
Recall details
- Recall number
- Z-0007-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 18, 2023
- Posted by the FDA
- Oct 1, 2024
- Recalling firm
- Philips North America
- Quantity
- 8,449 devices
- Product code
- DSI
- FDA submissions
- K151681
- Distribution
- US Nationwide. Global Distribution.
Product description
IntelliVue G7m Anesthesia Gas Module, Product Number 866173
Code information (models, lots, serial numbers)
UDI-DI: 00884838051492; Serial Numbers: DE43800129 through DE43812627
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IntelliVue G7m Gas Analyzer Module IntelliVue G7m Anesthesia Gas ModulePhilips Medizin Systeme Böblingen GmbH · ANALYZER, GAS, HALOTHANE, GASEOUS-PHASE (ANESTHETIC CONC.)1 recall2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗