Skip to content

CADD Medication Cassette Reservoir, 100mL, FS, Product Code… recallZ-0015-2025 · Smiths Medical ASD Inc. · initiated Jun 26, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

FDA root cause category: Process control

What the company is doing

Smiths Medical issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 06/26/2024 via mail. The notice explained the issue, potential risks, and requested the affected product be quarantined, and the response for returned to initiate the return and replacement process. Distributors were directed to notify their customers. For further inquires: Product complaints or adverse events: [email protected] 1-(866)-216-8806 Additional information or technical assistance 1-(800)-241-4002, option 3 [email protected]

Recall details

Recall number
Z-0015-2025
Status
Open, Classified
Initiated by the company
Jun 26, 2024
Posted by the FDA
Oct 18, 2024
Recalling firm
Smiths Medical ASD Inc.
Quantity
109563 units
Product code
FPA
FDA submissions
K081156
Distribution
Worldwide distribution.

Product description

CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JP

Code information (models, lots, serial numbers)

UID/DI NA, Lot/Serial Numbers: 3983315, 3971709, 3983316, 3983320, 3983329, 4013361, 4013363, 4013373, 4013377, 4013378, 4019427, 4019429, 4034028, 4037753, 4052410, 4040250, 4042843, 4048894, 4046846, 4052412, 4052413

Other devices under the same FDA submission

3 devices cleared under K081156 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95305).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page