Skip to content

The Trevo Trak 21 Microcatheter is a single-lumen, braided… recallZ-0021-2025 · Stryker Neurovascular · initiated Aug 30, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.

FDA root cause category: Under Investigation by firm

What the company is doing

On 09/12/2024, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL - Customer Notification" Letter to customers informing them that certain lots of Trevo Trak 21 Microcatheters were shipped with the incorrect configuration. The affected products contained Instructions for Use (IFUs) and product labels for international use, instead of the United States (US). The International IFUs contain intended uses that have not been cleared/approved for the U.S. market. Customer are instructed to: 1. Immediately check internal inventory for affected devices. 2. When using impacted lot(s) of Trevo Trak 21 Microcatheters, follow the US IFU attached in this communication NV00055463-02 (Attachment A) instead of the one that originally came with the device. 3. Circulate recall notification internally to all interested/affected parties and post this customer notification on or near affected products. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform Stryker Neurovascular if the subject devices have been distributed to other organizations and provide contact details so that Stryker Neurovascular can inform the recipients appropriately. 6. Inform Stryker Neurovascular of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916- 3876), Email: [email protected]). 7. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form allow for updating records and will also negate the need to send any further unnecessary communications on this matter. 8. Email the completed form to [email protected]. Contact the local Stryker representative, send an email to [email protected] or call US Customer Service: 1-855-91-NEURO (1-855-916-3876) for further assistance related to this Medical Device Recall Customer Notification.

Recall details

Recall number
Z-0021-2025
Status
Open, Classified
Initiated by the company
Aug 30, 2024
Posted by the FDA
Oct 7, 2024
Recalling firm
Stryker Neurovascular
Quantity
6
Product code
QJP
FDA submissions
K211594
Distribution
U.S. Nationwide distribution in the states of CA and VA.

Product description

The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. A rotating hemostatic valve with side-arm adapter is provided with each microcatheter.

Code information (models, lots, serial numbers)

Model/Catalog Number: 90338 UDI Codes: (01)07613327313956(17)251125(10)0000590444 (01)07613327313956(17)260101(10)0000603275 Lot Numbers: 0000590444 0000603275

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page