Why it was recalled

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Customer and Dealer notification will occur by means of letters alerting customers and dealers to the potential for overexposure to children if the conditions are not checked after the key switch is turned on. The notices also include the plan for the affected mobile x-ray systems. The system will be upgraded, free of charge, with updated software so that the exposure values set by the user are reflected regardless of the state of the key switch.

Recall details

Recall number
Z-3158-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Nov 23, 2023
Posted by the FDA
Sep 25, 2024
Recalling firm
FUJIFILM Healthcare Americas Corporation
Quantity
130
Product code
IZL
FDA submissions
K170858
Distribution
US Nationwide and Worldwide

Product description

The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Code information (models, lots, serial numbers)

All serial numbers shipped from FUJIFILM

Other devices under the same FDA submission

1 device cleared under K170858 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.