Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W… recallZ-0005-2025 · Zyno Medical LLC · initiated Sep 13, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zyno Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a defect in the air-in-line software algorithm.

FDA root cause category: Process change control

What the company is doing

An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 9/13/24 was sent to customers. Mitigation Zyno Medical LLC is mitigating this software issue by conducting a voluntary medical device correction to remove the faulty software and correctly install software that is free of this defect. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF Infusion Pump Air-in-Line Software Issue Response Verification Form. This will include checking your inventory for the affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form. A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action. Pass this notice to the appropriate personnel who need to be informed if the identified devices have been transferred to another location. If you have any questions or would like assistance, please contact your local Business Development Manager. Contact Information: [email protected] ***Update 1/3/2025***Additional consignees were sent URGENT: CLASS 1 RECALL letters due to additional affected units identified. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF I

Recall details

Recall number
Z-0005-2025
Status
Open, Classified
Initiated by the company
Sep 13, 2024
Posted by the FDA
Oct 9, 2024
Recalling firm
Zyno Medical LLC
Quantity
34,994 units (1819 units still need correction)
Product code
FRN
FDA submissions
K130690, K100705
Distribution
US Nationwide.

Product description

Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z

Code information (models, lots, serial numbers)

UDI-DI: Z-800 - 00814371020006 Z-800W - 00814371020020 Z-800F - 00814371020013 Z-800WF - 00814371020037 Lot Numbers: 190722360 20110921-SH 20111005-SH 20111102-SH 20111212-SH 20120105-SH 20120222-SH 20120501-SH 20120612-SH 20120719-SH 20120809-SH 20190809-SH 20120910-SH-1 20121213-sh 20130123-SH 20130227-SH 20120417-SH 20130511-SH 20130517-SH 20130609-SH 20130701-SH 20130812-SH 20131024-SH