Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W… recallZ-0005-2025 · Zyno Medical LLC · initiated Sep 13, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zyno Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a defect in the air-in-line software algorithm.
FDA root cause category: Process change control
What the company is doing
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 9/13/24 was sent to customers. Mitigation Zyno Medical LLC is mitigating this software issue by conducting a voluntary medical device correction to remove the faulty software and correctly install software that is free of this defect. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF Infusion Pump Air-in-Line Software Issue Response Verification Form. This will include checking your inventory for the affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form. A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action. Pass this notice to the appropriate personnel who need to be informed if the identified devices have been transferred to another location. If you have any questions or would like assistance, please contact your local Business Development Manager. Contact Information: [email protected] ***Update 1/3/2025***Additional consignees were sent URGENT: CLASS 1 RECALL letters due to additional affected units identified. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF I
Recall details
- Recall number
- Z-0005-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 13, 2024
- Posted by the FDA
- Oct 9, 2024
- Recalling firm
- Zyno Medical LLC
- Quantity
- 34,994 units (1819 units still need correction)
- Product code
- FRN
- FDA submissions
- K130690, K100705
- Distribution
- US Nationwide.
Product description
Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z
Code information (models, lots, serial numbers)
UDI-DI: Z-800 - 00814371020006 Z-800W - 00814371020020 Z-800F - 00814371020013 Z-800WF - 00814371020037 Lot Numbers: 190722360 20110921-SH 20111005-SH 20111102-SH 20111212-SH 20120105-SH 20120222-SH 20120501-SH 20120612-SH 20120719-SH 20120809-SH 20190809-SH 20120910-SH-1 20121213-sh 20130123-SH 20130227-SH 20120417-SH 20130511-SH 20130517-SH 20130609-SH 20130701-SH 20130812-SH 20131024-SH
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page