Trinias recallZ-0139-2025 · Shimadzu Corporation Medical Systems Division · initiated Sep 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shimadzu Corporation Medical Systems Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

FDA root cause category: Software Design Change

What the company is doing

On 9/6/24, recall notices were sent to customers who were asked to do the following: 1) Firmware for the system controller requires updating. The work to correct this issue will be performed by your Authorized Service provider. 2) Schedule time with your Distributer to have your system updated. 3) Complete and return the response form to [email protected] Questions can be directed to the firm's National Technical Support: Toll Free: (844)487-2767 Available 8:00 to 5:00 PM, Monday - Friday

Recall details

Recall number
Z-0139-2025
Status
Open, Classified
Initiated by the company
Sep 6, 2024
Posted by the FDA
Oct 22, 2024
Recalling firm
Shimadzu Corporation Medical Systems Division
Quantity
13
Product code
OWB
FDA submissions
K123508
Distribution
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

Product description

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.

Code information (models, lots, serial numbers)

UDI-DI: 04540217058327, 04540217049066, 04540217058143 Serial Numbers: 41EFCD7AC001, 41E58C773001, 41EFCD79C001, 41EFCD78B001, 41EFC73A9001, 41EFCD7D4001, 41E58C76B001, 41EFC73D9001, 41EFCD77C001, 41EFC73A5001, 41EFC73C5001, 41EFCD79A001, 41EFCD7A6001.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K123508 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page