BAUSCH and LOMB recallZ-0178-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Sep 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 09/05/2024 via email and USPS. The notice explained the issue, hazard, and requested the destruction of the affected product. Distributors were directed to notify their consignees.
Recall details
- Recall number
- Z-0178-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 4, 2024
- Posted by the FDA
- Oct 23, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 27,906 units
- Product code
- OJK
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Product description
Medline brand medical procedure convenience kits, labeled as: 1) EYE PACK, REF DYNJ19492L; 2) PHACO PACK, REF DYNJ22354L; 3) EYE PACK, REF DYNJ24573K; 4) EYE PACK, REF DYNJ30327A; 5) BASIC CATARACT PACK, REF DYNJ36868F; 6) BASIC RETINA PACK, REF DYNJ40811A; 7) OPHTHALMOLOGY PACK CKHA-LF, REF DYNJ41769A; 8) CATARACT PACK WRO-LF, REF DYNJ41861; 9) EYE PACK, REF DYNJ43911; 10) PHACO PACK, REF DYNJ44748I; 11) DR DRUMMOND EYE PACK, REF DYNJ45344G; 12) OPHTHALMIC PACK, REF DYNJ47275A; 13) CATARACT PACK, REF DYNJ49412B; 14) CATARACT PACK, REF DYNJ50948A; 15) CATARACT PACK, REF DYNJ50948A; 16) CATARACT DR MALONE PACK, REF DYNJ51590A; 17) CATARACT DR MALONE PACK, REF DYNJ51590A; 18) CUSTOM EYE TRAY, REF DYNJ51821B; 19) EYE SURGERY PACK, REF DYNJ53097C; 20) CATARACT KIT, REF DYNJ53944; 21) EYE PACK, REF DYNJ54847D; 22) PK, OPHTH-EYE, REF DYNJ55070B; 23) EYE PACK, REF DYNJ55320D; 24) PK, OPHTH-RETINA, REF DYNJ59253B; 25) RETINA PACK, REF DYNJ60515; 26) ASED KIT, REF DYNJ60745D; 27) CATARACT PACK, REF DYNJ61321B; 28) EYE PACK, REF DYNJ61862A; 29) CATARACT PACK DR FREDERICK, REF DYNJ63163; 30) CATARACT DR ESTOPINAL, REF DYNJ63164; 31) EYE CATARACT (BASIC EYE), REF DYNJ63272C; 32) EYE CATARACT (BASIC EYE), REF DYNJ63272C; 33) HMSL EYE PACK, REF DYNJ64537A; 34) EYE PACK, REF DYNJ65271C; 35) CMC ODS OPHTHALMIC PACK, REF DYNJ66832; 36) BASIC CATARACT PACK, REF DYNJ69165D; 37) BASIC CATARACT PACK, REF DYNJ69165D; 38) LIFESPAN VITRECTOMY PACK, REF DYNJ80082; 39) CATARACT PACK WITH NO TUB, REF DYNJ80232; 40) RETINA PACK, REF DYNJ80483; 41) RETINA PACK, REF DYNJ81640; 42) WESTFALL OCCUPLASTY, REF DYNJ81796A; 43) EYE PACK, REF DYNJ83799C; 44) OPHTHALMIC PACK, REF DYNJ85275; 45) CATARACT, REF DYNJ908059A; 46) EYE TRAY 1-LF, REF LLSCAEYE01A; 47) PHACO PACK, REF LYN011PHNHD; 48) PHACO PACK, REF LYN011PHNHD; 49) SPECIALTY BESSER PACK, REF LYN023BPSSD; 50) SPECIALTY BESSER PACK, REF LYN023BPSSD; 51) SPECIALTY BESSER PACK, REF LYN023BPSSD; 52) BASIC OPHTHALMIC PACK, REF LYN023BSBLB; 53) BASIC OPHTHALMIC PACK, REF LYN023BSBLB; 54) OPHTHALMIC PACK, REF LYN023OPSOB; 55) CATARACT PACK CLEARSIGHT CTR, REF LYN024CACSB; 56) BASIC CATARACT PACK, REF LYN030BCSDB; 57) CATARACT PACK SEVEN SPRINGS, REF LYN030CPSSA; 58) AIKEN CATARACT PACK II, REF LYN030CTAKG; 59) AIKEN CATARACT PACK II, REF LYN030CTAKG; 60) AIKEN CATARACT PACK II, REF LYN030CTAKG; 61) VA EYE II-LF, REF LYN030VAEYA; 62) VA EYE II-LF, REF LYN030VAEYA; 63) CATARACT PACK, REF LYN035CTSEC; 64) EYES OF YORK DR BENZ CATARACT, REF LYN037DBEYM; 65) EYES OF YORK DR BENZ CATARACT, REF LYN037DBEYM; 66) OPHTHALMIC PACK, REF LYN041OPFCC; 67) OPHTHALMIC PACK, REF LYN041OPFCC; 68) OPHTHALMIC PACK, REF LYN041OPFCC; 69) OPHTHALMIC PACK, REF LYN041OPFCC; 70) CATARACT PACK, REF LYN042CPFEG; 71) OPTHALMIC PACK-LF, REF PHS529724000B; 72) CATARACT PACK, REF VAL001CPMEB; 73) CATARACT PACK, REF VAL001CPMEB; 74) THE SURGERY CENTER CATARACT, REF VAL002CATSG; 75) THE SURGERY CENTER CATARACT, REF VAL002CATSG; 76) RETINA PACK PIEDMONT EYE, REF VAL003REPEA; 77) CATARACT PACK, REF VAL009CAAHC; 78) RENAISSANCE CAT PACK, REF VAL009CAREA; 79) CATARACT TENAYA SURGERY CENTER, REF VAL009CATEA; 80) CATARACT PACK, REF VAL009CTSGA; 81) DOUGHERTY LASER VISION- NEW II, REF VAL009EYDLC; 82) RETINA PACK-TRINITY GROUP, REF VAL009RETGA; 83) RENAISSANCE VIT PACK, REF VAL009VIREA; 84) CATARACT PACK APERTURE, REF VAL011CTVAA; 85) OPHTHALMIC PACK, REF VAL022BSOMD; 86) CATARACT PACK - INST OF OKLA, REF VAL024CAIOD; 87) EYE PACK, REF VAL024EYTRB; 88) CATARACT PACK, REF VAL026CTNFC; 89) EYE PACK, REF VAL029EYGSB; 90) INTRAOCULAR PACK, REF VAL031IOGCB; 91) CATARACT, REF VAL035CPAEC; 92) CATARACT PACK 2, REF VAL036CABRD; 93) BRINTON VISION ICL PACK, REF VAL037ICBVG; 94) EYE PACK, REF VAL038BSCID; 95) CATARACT PACK EYECARE CONSULT, REF VAL038CTECB; 96) EYE PACK SC OF SOUTHWOODS, REF VAL041EPSWA; 97) BOOZMAN HOF PACK 2, REF VAL042EYBHC; 98) CATARACT PACK WILD
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
1) REF DYNJ19492L, UDI/DI 40195327368587 (each), 10195327368586 (case), Lot Numbers: 24ABE741; 2) REF DYNJ22354L, UDI/DI 40195327239924 (each), 10195327239923 (case), Lot Numbers: 23LBN139; 3) REF DYNJ24573K, UDI/DI 40195327144303 (each), 10195327144302 (case), Lot Numbers: 23KBB931; 4) REF DYNJ30327A, UDI/DI 40193489997089 (each), 10193489997088 (case), Lot Numbers: 24ABB714; 5) REF DYNJ36868F, UDI/DI 40888277897060 (each), 10888277897069 (case), Lot Numbers: 23KBD160; 6) REF DYNJ40811A, UDI/DI 40888277014634 (each), 10888277014633 (case), Lot Numbers: 23KBE584; 7) REF DYNJ41769A, UDI/DI 40888277267825 (each), 10888277267824 (case), Lot Numbers: 23KBB744; 8) REF DYNJ41861, UDI/DI 40888277020741 (each), 10888277020740 (case), Lot Numbers: 23KBB746; 9) REF DYNJ43911, UDI/DI 40888277073617 (each), 10888277073616 (case), Lot Numbers: 23LBQ741; 10) REF DYNJ44748I, UDI/DI 40193489273459 (each), 10193489273458 (case), Lot Numbers: 23LBP074; 11) REF DYNJ45344G, UDI/DI 40195327402922 (each), 10195327402921 (case), Lot Numbers: 23LBU068; 12) REF DYNJ47275A, UDI/DI 40888277571946 (each), 10888277571945 (case), Lot Numbers: 24ABE470; 13) REF DYNJ49412B, UDI/DI 40889942935278 (each), 10889942935277 (case), Lot Numbers: 24ABD256; 14) REF DYNJ50948A, UDI/DI 40195327117550 (each), 10195327117559 (case), Lot Numbers: 24ABC786; 15) REF DYNJ50948A, UDI/DI 40195327117550 (each), 10195327117559 (case), Lot Numbers: 23KBA350; 16) REF DYNJ51590A, UDI/DI 40193489721387 (each), 10193489721386 (case), Lot Numbers: 23LBN087; 17) REF DYNJ51590A, UDI/DI 40193489721387 (each), 10193489721386 (case), Lot Numbers: 23KBE531; 18) REF DYNJ51821B, UDI/DI 40193489963374 (each), 10193489963373 (case), Lot Numbers: 23KBE568; 19) REF DYNJ53097C, UDI/DI 40889942721376 (each), 10889942721375 (case), Lot Numbers: 24ABC180; 20) REF DYNJ53944, UDI/DI 40889942450689 (each), 10889942450688 (case), Lot Numbers: 23LBR325; 21) REF DYNJ54847D, UDI/DI 40193489266659 (each), 10193489266658 (case), Lot Numbers: 23LBI510; 22) REF DYNJ55070B, UDI/DI 40193489800686 (each), 10193489800685 (case), Lot Numbers: 23LBU001; 23) REF DYNJ55320D, UDI/DI 40195327429219 (each), 10195327429218 (case), Lot Numbers: 24ABD339; 24) REF DYNJ59253B, UDI/DI 40193489269322 (each), 10193489269321 (case), Lot Numbers: 24ABB328; 25) REF DYNJ60515, UDI/DI 40193489510240 (each), 10193489510249 (case), Lot Numbers: 24ABE890; 26) REF DYNJ60745D, UDI/DI 40193489949309 (each), 10193489949308 (case), Lot Numbers: 24ABA754; 27) REF DYNJ61321B, UDI/DI 40193489794411 (each), 10193489794410 (case), Lot Numbers: 23KBF225; 28) REF DYNJ61862A, UDI/DI 40195327216826 (each), 10195327216825 (case), Lot Numbers: 23KBC772; 29) REF DYNJ63163, UDI/DI 40193489721585 (each), 10193489721584 (case), Lot Numbers: 23LBL330; 30) REF DYNJ63164, UDI/DI 40193489721592 (each), 10193489721591 (case), Lot Numbers: 23LBN120; 31) REF DYNJ63272C, UDI/DI 40195327532254 (each), 10195327532253 (case), Lot Numbers: 24ABC862; 32) REF DYNJ63272C, UDI/DI 40195327532254 (each), 10195327532253 (case), Lot Numbers: 23KBI662; 33) REF DYNJ64537A, UDI/DI 40195327518494 (each), 10195327518493 (case), Lot Numbers: 24ABC540; 34) REF DYNJ65271C, UDI/DI 40193489957946 (each), 10193489957945 (case), Lot Numbers: 23KBC696; 35) REF DYNJ66832, UDI/DI 40193489389051 (each), 10193489389050 (case), Lot Numbers: 23KBE083; 36) REF DYNJ69165D, UDI/DI 40195327224340 (each), 10195327224349 (case), Lot Numbers: 23LBP941; 37) REF DYNJ69165D, UDI/DI 40195327224340 (each), 10195327224349 (case), Lot Numbers: 23KBE571; 38) REF DYNJ80082, UDI/DI 40195327012008 (each), 10195327012007 (case), Lot Numbers: 24ABD342; 39) REF DYNJ80232, UDI/DI 40195327017065 (each), 10195327017064 (case), Lot Numbers: 23LBN193; 40) REF DYNJ80483, UDI/DI 40195327045341 (each), 10195327045340 (case), Lot Numbers: 24ABC583; 41) REF DYNJ81640, UDI/DI 40195327129478 (each), 10195327129477 (case), Lot Numbers: 23LBK169; 42) REF DYNJ81796A, UDI/DI 40195327570386 (each), 10195327570385 (case), Lot Numbers: 23LBN670; 43) REF DYNJ83799C, UDI/DI 40195327443932 (each), 10195327443931 (case), Lot Numbers: 23LBP953; 44) REF DYNJ85275, UDI/DI 40195327484690 (each), 10195327484699 (case), Lot Numbers: 24ABA128; 45) REF DYNJ908059A, UDI/DI 40195327125432 (each), 10195327125431 (case), Lot Numbers: 23LBT833; 46) REF LLSCAEYE01A, UDI/DI 40884389376981 (each), 10884389376980 (case), Lot Numbers: 23KBB457; 47) REF LYN011PHNHD, UDI/DI 40193489234405 (each), 10193489234404 (case), Lot Numbers: 24ABD326; 48) REF LYN011PHNHD, UDI/DI 40193489234405 (each), 10193489234404 (case), Lot Numbers: 24ABA918; 49) REF LYN023BPSSD, UDI/DI 40193489640954 (each), 10193489640953 (case), Lot Numbers: 24ABA470; 50) REF LYN023BPSSD, UDI/DI 40193489640954 (each), 10193489640953 (case), Lot Numbers: 23LBN465; 51) REF LYN023BPSSD, UDI/DI 40193489640954 (each), 10193489640953 (case), Lot Numbers: 23KBF760; 52) REF LYN023BSBLB, UDI/DI 40889942336242 (each), 10889942336241 (case), Lot Numbers: 24ABC586; 53) REF LYN023BSBLB, UDI/DI 40889942336242 (each), 10889942336241 (case), Lot Numbers: 24ABD327; 54) REF LYN023OPSOB, UDI/DI 40889942375791 (each), 10889942375790 (case), Lot Numbers: 24ABD328; 55) REF LYN024CACSB, UDI/DI 40195327095735 (each), 10195327095734 (case), Lot Numbers: 23LBN673; 56) REF LYN030BCSDB, UDI/DI 40193489457965 (each), 10193489457964 (case), Lot Numbers: 24ABH583; 57) REF LYN030CPSSA, UDI/DI 40195327128549 (each), 10195327128548 (case), Lot Numbers: 23KBD385; 58) REF LYN030CTAKG, UDI/DI 40193489465946 (each), 10193489465945 (case), Lot Numbers: 24ABA502; 59) REF LYN030CTAKG, UDI/DI 40193489465946 (each), 10193489465945 (case), Lot Numbers: 23LBP443; 60) REF LYN030CTAKG, UDI/DI 40193489465946 (each), 10193489465945 (case), Lot Numbers: 23LBQ089; 61) REF LYN030VAEYA, UDI/DI 40195327242917 (each), 10195327242916 (case), Lot Numbers: 24ABA893; 62) REF LYN030VAEYA, UDI/DI 40195327242917 (each), 10195327242916 (case), Lot Numbers: 24ABD329; 63) REF LYN035CTSEC, UDI/DI 40889942532026 (each), 10889942532025 (case), Lot Numbers: 23KBC621; 64) REF LYN037DBEYM, UDI/DI 40193489473910 (each), 10193489473919 (case), Lot Numbers: 23LBQ090; 65) REF LYN037DBEYM, UDI/DI 40193489473910 (each), 10193489473919 (case), Lot Numbers: 23KBD746; 66) REF LYN041OPFCC, UDI/DI 40889942357841 (each), 10889942357840 (case), Lot Numbers: 24ABD867; 67) REF LYN041OPFCC, UDI/DI 40889942357841 (each), 10889942357840 (case), Lot Numbers: 23LBN722; 68) REF LYN041OPFCC, UDI/DI 40889942357841 (each), 10889942357840 (case), Lot Numbers: 23LBA842; 69) REF LYN041OPFCC, UDI/DI 40889942357841 (each), 10889942357840 (case), Lot Numbers: 23KBD251; 70) REF LYN042CPFEG, UDI/DI 40193489219372 (each), 10193489219371 (case), Lot Numbers: 23KBC622; 71) REF PHS529724000B, UDI/DI 40195327166022 (each), 10195327166021 (case), Lot Numbers: 23KBE642; 72) REF VAL001CPMEB, UDI/DI 40195327205455 (each), 10195327205454 (case), Lot Numbers: 24ABE054; 73) REF VAL001CPMEB, UDI/DI 40195327205455 (each), 10195327205454 (case), Lot Numbers: 23LBH892; 74) REF VAL002CATSG, UDI/DI 40195327250417 (each), 10195327250416 (case), Lot Numbers: 24ABN974; 75) REF VAL002CATSG, UDI/DI 40195327250417 (each), 10195327250416 (case), Lot Numbers: 23KBC653; 76) REF VAL003REPEA, UDI/DI 40889942968382 (each), 10889942968381 (case), Lot Numbers: 24ABF774; 77) REF VAL009CAAHC, UDI/DI 40195327260300 (each), 10195327260309 (case), Lot Numbers: 23KBB402; 78) REF VAL009CAREA, UDI/DI 40193489462242 (each), 10193489462241 (case), Lot Numbers: 23KBC624; 79) REF VAL009CATEA, UDI/DI 40889942872399 (each), 10889942872398 (case), Lot Numbers: 24ABA924; 80) REF VAL009CTSGA, UDI/DI 40193489577540 (each), 10193489577549 (case), Lot Numbers: 24ABA478; 81) REF VAL009EYDLC, UDI/DI 40195327302291 (each), 10195327302290 (case), Lot Numbers: 24ABE052; 82) REF VAL009RETGA, UDI/DI 40193489676991 (each), 10193489676990 (case), Lot Numbers: 23LBQ654; 83) REF VAL009VIREA, UDI/DI 40193489462259 (each), 10193489462258 (case), Lot Numbers: 24ABA925; 84) REF VAL011CTVAA, UDI/DI 40889942005568 (each), 10889942005567 (case), Lot Numbers: 23KBC625; 85) REF VAL022BSOMD, UDI/DI 40889942760214 (each), 10889942760213 (case), Lot Numbers: 23LBQ097; 86) REF VAL024CAIOD, UDI/DI 40193489491525 (each), 10193489491524 (case), Lot Numbers: 23KBB700; 87) REF VAL024EYTRB, UDI/DI 40193489426008 (each), 10193489426007 (case), Lot Numbers: 23LBN467; 88) REF VAL026CTNFC, UDI/DI 40193489458412 (each), 10193489458411 (case), Lot Numbers: 23LBA843; 89) REF VAL029EYGSB, UDI/DI 40193489550727 (each), 10193489550726 (case), Lot Numbers: 23KBT152; 90) REF VAL031IOGCB, UDI/DI 40889942307198 (each), 10889942307197 (case), Lot Numbers: 24ABA898; 91) REF VAL035CPAEC, UDI/DI 40889942537373 (each), 10889942537372 (case), Lot Numbers: 23KBE643; 92) REF VAL036CABRD, UDI/DI 40195327196159 (each), 10195327196158 (case), Lot Numbers: 23KBC657; 93) REF VAL037ICBVG, UDI/DI 40193489868136 (each), 10193489868135 (case), Lot Numbers: 23KBC628; 94) REF VAL038BSCID, UDI/DI 40888277761767 (each), 10888277761766 (case), Lot Numbers: 23KBC629; 95) REF VAL038CTECB, UDI/DI 40889942867883 (each), 10889942867882 (case), Lot Numbers: 23LBP453; 96) REF VAL041EPSWA, UDI/DI 40193489394550 (each), 10193489394559 (case), Lot Numbers: 24ABA469; 97) REF VAL042EYBHC, UDI/DI 40195327033249 (each), 10195327033248 (case), Lot Numbers: 23KBD595; 98) REF VAL046CAWCA, UDI/DI 40193489846721 (each), 10193489846720 (case), Lot Numbers: 23LBM013; 99) REF VAL047VIVOA, UDI/DI 40193489375122 (each), 10193489375121 (case), Lot Numbers: 23LBM392; 100) REF VAL048TSCCA, UDI/DI 40195327250455 (each), 10195327250454 (case), Lot Numbers: 23LBQ099; 101) REF VAL048TSCCA, UDI/DI 40195327250455 (each), 10195327250454 (case), Lot Numbers: 23KBC660; 102) REF VAL050CTFEI, UDI/DI 40193489368063 (each), 10193489368062 (case), Lot Numbers: 24ABA899; 103) REF VAL050EEPKB, UDI/DI 40195327267149 (each), 10195327267148 (case), Lot Numbers: 24ABA472; 104) REF VAL050ITKEA, UDI/DI 40195327134090 (each), 10195327134099 (case), Lot Numbers: 23KBK068; 105) REF VAL050REBOB, UDI/DI 40193489833653 (each), 10193489833652 (case), Lot Numbers: 24ABA477; 106) REF VAL050VIAHD, UDI/DI 40195327195565 (each), 10195327195564 (case), Lot Numbers: 23JBU308; 107) REF VAL054CAMAA, UDI/DI 40193489228114 (each), 10193489228113 (case), Lot Numbers: 23LBQ100; 108) REF VAL059DKSWB, UDI/DI 40193489667951 (each), 10193489667950 (case), Lot Numbers: 23KBE384; 109) REF VAL075EYLSA, UDI/DI 40889942804345 (each), 10889942804344 (case), Lot Numbers: 23LBP466; 110) REF VAL089EYRUA, UDI/DI 40193489438162 (each), 10193489438161 (case), Lot Numbers: 23LBP468
Devices named in this recall
92 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ44748IMedline Industries, Inc. · Eye tray2 recalls0 reports
- Medline Industries, Inc. VAL009CTSGAMedline Industries, Inc. · Eye tray2 recalls0 reports
- Medline Industries, Inc. DYNJ80082Medline Industries, Inc. · Eye tray2 recalls0 reports
- Medline Industries, Inc. VAL050EEPKBMedline Industries, Inc. · Eye tray2 recalls0 reports
- Medline Industries, Inc. LYN023BSBLBMedline Industries, Inc. · Eye tray2 recalls0 reports
- Medline Industries, Inc. VAL002CATSGMedline Industries, Inc. · Eye tray1 recall5 reports
- Medline Industries, Inc. VAL001CPMEBMedline Industries, Inc. · Eye tray1 recall2 reports
- Medline Industries, Inc. LYN030BCSDBMedline Industries, Inc. · Eye tray1 recall1 report
- Medline Industries, Inc. VAL048TSCCAMedline Industries, Inc. · Eye tray1 recall1 report
- Medline Industries, Inc. LYN041OPFCCMedline Industries, Inc. · Eye tray1 recall1 report
- Medline Industries, Inc. VAL009CATEAMedline Industries, Inc. · Eye tray1 recall1 report
- BAUSCH and LOMB LYN042CPFEGMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL054CAMAAMedline Industries, Inc. · Eye tray1 recall0 reports
- BAUSCH and LOMB LYN011PHNHDMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ54847DMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ59253BMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL050CTFEIMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL047VIVOAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ66832Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL041EPSWAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL024EYTRBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL089EYRUAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL026CTNFCMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL009CAREAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL009VIREAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN030CTAKGMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN037DBEYMMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL024CAIODMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ60515Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL029EYGSBMedline Industries, Inc. · Eye tray1 recall0 reports
- BAUSCH and LOMB LYN023BPSSDMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL059DKSWBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL009RETGAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ51590AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ63163Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ63164Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ61321BMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ55070BMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL050REBOBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL046CAWCAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL037ICBVGMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ60745DMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ65271CMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ51821BMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ30327AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ80232Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL042EYBHCMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ80483Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN024CACSBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ50948AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ908059AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN030CPSSAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ81640Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL050ITKEAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ24573KMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. PHS529724000BMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL050VIAHDMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL036CABRDMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ61862AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ69165DMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ22354LMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN030VAEYAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL009CAAHCMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL009EYDLCMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ19492LMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ45344GMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ55320DMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ83799CMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ85275Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ64537AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ63272CMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ81796AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LLSCAEYE01AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ40811AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ41861Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ43911Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ41769AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ47275AMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL038BSCIDMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ36868FMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL011CTVAAMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL031IOGCBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN023OPSOBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ53944Medline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. LYN035CTSECMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL035CPAECMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ53097CMedline Industries, Inc. · Eye tray1 recall0 reports
- BAUSCH and LOMB VAL022BSOMDMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL075EYLSAMedline Industries, Inc. · Eye tray1 recall0 reports
- BAUSCH and LOMB VAL038CTECBMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. DYNJ49412BMedline Industries, Inc. · Eye tray1 recall0 reports
- Medline Industries, Inc. VAL003REPEAMedline Industries, Inc. · Eye tray1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95415).
- Z-0148-2025 · Open, ClassifiedCenturion PEDIATRIC/ADULT TRACHEOSTOMY T, REF TC7895; medical procedure convenience kit
- Z-0149-2025 · Open, ClassifiedCenturion EP LAB INSTRUMENT KIT, REF MNS10480; medical procedure convenience kit
- Z-0150-2025 · Open, ClassifiedCenturion brand, medical procedure convenience kits, labeled as: a) FLOOR - UMBILICAL CUT DOWN TRA, b) PLASTICS SUTURE TRAY
- Z-0151-2025 · Open, ClassifiedCenturion OB TRAY, REF MNS13965A; medical procedure convenience kit
- Z-0152-2025 · Open, ClassifiedMedline CV ANESTHESIA - ROOM SET UP, REF DYNJ905503F; medical procedure convenience kit
- Z-0153-2025 · Open, ClassifiedMedline brand medical procedure convenience kit, labeled as: a) PAIN PROCEDURE PACK, REF DYNJ80370A; b) DR HEFZY BLOCK TRAY, REF DYNJRA1854A
- Z-0154-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: a) SET UP PACK, REF DYNJ36320A; b) ANGIOGRAM WRO-LF, REF DYNJ41876B
- Z-0155-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: a) RADIOLOGY LP TRAY, REF DYNDH1609; b) CARDIAC CATH PACK-LF, REF DYNJ19010I; c) SUPPLEMENTAL NEURO ANGIO PK, REF…
- Z-0156-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: a) CUH OPEN HEART A B C - CDS, REF CDS983202I; a) PICC INSERTION KIT, REF DYNDC2486A; a) OPEN HEART PACK, REF DYN…
- Z-0157-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) KIT LINE INSERTION, REF CVI4730A; b) KIT LINE INSERTION, REF CVI4730A; c) LINE INSERTION TRAY, REF DYNJ20803D; …
- Z-0158-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) CONE BIOPSY PACK, REF DYNJ17534L; b) BIOPSY TRAY, REF DYNJ43076A; c) LOU LOOP RECORDER PACK, REF DYNJ49404C; d…
- Z-0159-2025 · Open, ClassifiedMedline DENTAL PACK, REF DYNJ64086B; medical procedure convenience kit
- Z-0160-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: 1) T&A CDS-LF, REF CDS982091I; 2) TONSIL CDS, REF CDS984492B; 3) RV T&A PACK-LF, REF DYNJ0798129I; 4) DIRECT MICR…
- Z-0161-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: 1) CYSTO CDS, REF CDS983525D; 2) CYSTO CDS, REF CDS983525D; 3) CYSTO PACK-LF, REF DYNJ25834B; 4) CYSTO PACK, REF …
- Z-0162-2025 · Open, ClassifiedMedline CENTRAL CATHETER INSERTION TRA, REF CVI680C; medical procedure convenience kit
- Z-0163-2025 · Open, ClassifiedMedline DIALYSIS PACK, REF DYNJ36932A; medical procedure convenience kit
- Z-0164-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) TRAY ADD A CATH FOL UM MF 10ML, REF DYND150200S; b) TRAY FOL LATX UM COUD 16FR MF, REF DYND150216CS; c) TRAY F…
- Z-0165-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) PLASTIC PACK, REF DYNJ42735A; b) DR SELEM PACK, REF DYNJ43457; c) BREAST PK TEGH-LF, REF DYNJ46717A; d) BREAST…
- Z-0166-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: 1) THORACENTESIS TRAY, REF 00-400616J; 2) HH PACK SONICATION SPEC CONTAI, REF 3706WHST; 3) T&A CDS, REF CDS980427J; …
- Z-0167-2025 · Open, ClassifiedMedline CHEST TUBE INSERTION TRAY ADUL, REF CHT1165; medical procedure convenience kit
- Z-0168-2025 · Open, ClassifiedMedline IV TEAM PICC, REF DYNJ48222; medical procedure convenience kit
- Z-0169-2025 · Open, ClassifiedMedline PEDIATRIC DRAPE PK, REF DYNJ81297; medical procedure convenience kit
- Z-0170-2025 · Open, ClassifiedMedline TUBAL LIGATION PACK, REF DYNJ66230B; medical procedure convenience kit
- Z-0171-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) UNIVERSAL DRAPE PACK, REF DYNJ44114; b) DRAPE PACK, REF DYNJ82663B
- Z-0173-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) NEURO PACK, REF DYNJ37909A; b) SHUNT PACK, REF DYNJ40040C
- Z-0174-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: 1) L&D/OB PACK-LF, REF DYNJ0213264O; 2) VAG DELIVERY HLTH ALLIA PK-LF, REF DYNJ0375486K; 3) VAG DELIVERY HLTH ALLIA…
- Z-0175-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) LITHOTOMY PACK, REF DYNJ24362C; b) MAJOR GYNE LAPAROSCOPY SHSC, REF DYNJ46774B; c) GYN ANCILARY PACK, REF DYNJ8…
- Z-0176-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) ABDOMINAL HYSTERECTOMY TEGH-LF, REF DYNJ46730; b) HYSTEROSCOPY PACK, REF DYNJ49302A; c) PERI GYN PACK, REF DYNJ…
- Z-0177-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: a) OB PACK-LF, REF DYNJ0912449M; b) OB PACK, REF DYNJ17309G; c) OB PK, REF DYNJ34268D; d) OB DELIVERY PACK, REF D…
- Z-0179-2025 · Open, ClassifiedMedline brand medical procedure convenience kits, labeled as: 1) NEW LONDON TOTAL KNEE CDS, REF CDS982519W; 2) HAND/FOOT CDS, REF CDS984003A; 3) SPECIAL PROCEDURE PACK, REF DYN…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page