SYNCHROMED II recallZ-0137-2025 · Medtronic Neuromodulation · initiated Sep 11, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for error and inability to interrogate pump due to data corruption in the pump memory.

FDA root cause category: Device Design

What the company is doing

On 11-September-2024, a Medtronic representative was sent to the physician s facility where the representative provided the communication and confirmation form to the physician. The physician was asked to sign the confirmation form indicating they read the notification. The representative reset the patient s pump.

Recall details

Recall number
Z-0137-2025
Status
Open, Classified
Initiated by the company
Sep 11, 2024
Posted by the FDA
Oct 21, 2024
Recalling firm
Medtronic Neuromodulation
Quantity
1 unit
Product code
LKK
Distribution
US Nationwide distribution in the state of Mississippi.

Product description

SynchroMed II Infusion Pump. Product Number: 8637-20

Code information (models, lots, serial numbers)

GTIN Number: 00643169630505. Serial Number: NGP004814H

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates