Skip to content

Product Name: Insight Base Flex - 8, Ref 608-3TX recallZ-0260-2025 · Trackx Technology Inc · initiated Sep 3, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Trackx Technology Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-01. This is reflected on the carton label and sterile pouch label correctly.

FDA root cause category: Error in labeling

What the company is doing

TrackX Technology issued an Urgent: Medical Device Field Correction notice to its consignees via hand delivery by their sales representatives on 10/01/2024. The notice explained the issue and requested that the notice be provided to all parties to whom the product was distributed. The sales representatives exchanged the affected product for correctly labeled product.

Recall details

Recall number
Z-0260-2025
Status
Open, Classified
Initiated by the company
Sep 3, 2024
Posted by the FDA
Oct 30, 2024
Recalling firm
Trackx Technology Inc
Quantity
82
Product code
OWB
FDA submissions
K173736, K200360
Distribution
NC, VA, FL, PA, NY

Product description

Product Name: Insight Base Flex - 8, Ref 608-3TX; The device consists of a sterile Snap that includes reflective features that facilitate imaging of the (attached) instrument using instrument tracking technology when used with a compatible tracking system. Each TrackX Snap has a unique set of 2 or more reflective markers. By having unique configurations of reflective markers, it is possible to identify which TrackX Snap is in the field of view of the camera(s). As such, it is also possible to have multiple TrackX Snaps in the field at once and to have different TrackX Snaps for different surgical instruments. The device is considered to be a sterile accessory.

Code information (models, lots, serial numbers)

GTIN: 00857088008493, Lot: 2404002, Expiration Date: 04/01/2027

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates