Skip to content

17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay… recallZ-0258-2025 · DRG International, Inc. · initiated Sep 13, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DRG International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

FDA root cause category: Process change control

What the company is doing

DRG issued Urgent Field Safety Notification on 9/17/24 via email. Letter states reason for recall, health risk and action to take: Please check your inventory for the respective lot and ensure that the proper acceptance ranges provided below are used to validate the assay. Immediately reply to email: [email protected] to confirm this notice and your compliance.

Recall details

Recall number
Z-0258-2025
Status
Open, Classified
Initiated by the company
Sep 13, 2024
Posted by the FDA
Oct 29, 2024
Recalling firm
DRG International, Inc.
Quantity
504 units
Product code
JLX
Distribution
AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam

Product description

17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma Model/Catalog Number: EIA-1292

Code information (models, lots, serial numbers)

Lot Code: 04048474012927 (DRG Instruments GmbH) lots: 22K024 and 22K034

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page