bk3000, model #2300-11 recallZ-0136-2025 · B-K Medical A/S · initiated Sep 4, 2024

Class II Open, Classified All recalls posted in October 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from B-K Medical A/S or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

FDA root cause category: Process control

What the company is doing

Notifications were issued on and dated 9/4/2024 informing the consignee of the safety issue and the actions to be taken, which included: Until GE HealthCare implements a permanent correction on the device for this issue, the consignee is to inspect the power cord connection at the power inlet to the ultrasound device. A photograph of the area to be inspected was embedded in the letter. If a cord retaining bracket is not present, the consignee is to ensure the power cord is fully pushed into the power inlet before each use and while the ultrasound device is charging. If a cord retaining bracket is present, no further action is required. The consignee is to ensure all potential staff in the facility are made aware of the safety notification and the recommended actions. An acknowledgment form was enclosed for completion and return.

Recall details

Recall number
Z-0136-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 4, 2024
Posted by the FDA
Oct 18, 2024
Recalling firm
B-K Medical A/S
Quantity
1648 devices
Product code
IYN
FDA submissions
K223830
Distribution
Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.

Product description

bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option.

Code information (models, lots, serial numbers)

All serial numbers, GTIN 05704916000264.