Surg Strp 2x6 recallZ-0254-2025 · Integra LifeSciences Corp. · initiated Sep 20, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
FDA root cause category: Nonconforming Material/Component
What the company is doing
Integra issued 1st Notification- Urgent: Voluntary Medical Device Recall letter via E-mail and FedEx overnight mail on 9/20/24. Letter states reason for recall, health risk and action to take: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service and quarantine them. 2. If you do have affected product, check the box on the enclosed Acknowledgement Form I do have affected product. Record the lot number and total quantity of the affected product that you have. 3. Forward this notice to those who utilize the product so they are aware of this recall and can identify and quarantine any affected product that may remain in clinical areas. 4. If you do not have affected product, check the box, I do not have affected product. 5. Complete the rest of the Acknowledgement Form and return to [email protected] or FAX to 1-609-750-4220. 6. Keep a copy of the form for your records. 7. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. You can request a credit for the quantities which you returned. PLEASE NOTE THAT REGARDLESS OF WHETHER YOU HAVE THE AFFECTED PRODUCT TO RETURN OR NOT A COMPLETED ACKNOWLEDGEMENT IS REQUIRED Additional actions to be Taken by Integra Sales Representatives and/or Distributors: 7. Check your customer traceability records for shipments of above catalog and lot numbers. 8. Forward a copy of the enclosed Field Safety Notice to any of your customers that have purchased the above catalog and lot number. 9. Distributors can request a credit for the quantities which they returned. Should you have any questions regarding these instructions, please contact your Integra Sales Representative or Customer Service from Monday to Friday 8:00 a.m. 8:00 p.m. EST at 1-800-654-2873 or [email protected].
Recall details
- Recall number
- Z-0254-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 20, 2024
- Posted by the FDA
- Oct 29, 2024
- Recalling firm
- Integra LifeSciences Corp.
- Quantity
- 1222
- Product code
- HBA
- FDA submissions
- K193346
- Distribution
- Worldwide Distribution. US Nationwide, Canada Germany United Kingdom of Great Britain and Northern Ireland (the) Malta Italy Barbados Trinidad and Tobago Australia New Zealand Netherlands (the) Belgium France Ireland Switzerland Brunei Darussalam Bahamas (the) South Africa China Ecuador Korea (the Republic of) Czechia Kuwait Luxembourg French Guiana Mayotte Burundi Reunion Martinique Poland Costa Rica Japan Senegal Turkey Serbia Mexico Cambodia Spain Ghana Iceland Jersey Iraq Bahrain Lebanon Dominican Republic (the) Singapore Portugal Greece Austria C¿te d'Ivoire Palestine, State of Georgia Nigeria Slovenia Sweden Colombia United Arab Emirates (the) Saudi Arabia Qatar Hong Kong Oman Albania Panama Holy See (the) Kenya Pakistan Guernsey Morocco Chile Brazil Argentina Puerto Rico Rwanda Jordan Philippines (the) Hungary Romania Bolivia (Plurinational State of) Malaysia El Salvador Peru Lithuania Republic of North Macedonia Taiwan (Province of China) Israel Cyprus Bulgaria
Product description
SURG STRP 2X6 Model/Catalog Number: 801455. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Code information (models, lots, serial numbers)
UDI-DI: 10381780515111, 20886704036637 Note: the same product code was distributed for a few years under JnJ under a transition agreement with Integra. Therefore, 2 UDI numbers are listed here for JnJ and for Integra. All unexpired products.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- CODMAN® 801455Integra Lifesciences Production Corporation · Neurosurgical paddie1 recall1 report
- CODMAN 80-1455Cerenovus, Inc. · NEUROSURGICAL PADDIE1 recall0 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K193346 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CODMAN® 245404Integra Lifesciences Production Corporation · Neurosurgical paddie10 reports
- CODMAN® 245431Integra Lifesciences Production Corporation · Neurosurgical paddie2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95432).
- Z-0235-2025 · Open, ClassifiedMICR PATIE RND Model/Catalog Number: 801396. CODMAN Surgical Patties and CODMAN Surgical Strips are manufactured of COTTONOID Material with x-ray detectable markers. All patties ha…
- Z-0236-2025 · Open, ClassifiedSURG PAT XRAY 1/4X1-1/2 Model/Catalog Number: 801397. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues …
- Z-0237-2025 · Open, ClassifiedSURG PAT XRAY 1/4X3 Model/Catalog Number: 801398. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of t…
- Z-0238-2025 · Open, ClassifiedSURG PAT XRAY 1/4X1/4 Model/Catalog Number: 801399. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of…
- Z-0239-2025 · Open, ClassifiedSURG PAT XRAY 1/2X1/2 Model/Catalog Number: 801400. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of…
- Z-0240-2025 · Open, ClassifiedSURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of…
- Z-0241-2025 · Open, ClassifiedSURG PAT XRAY 1/2X1 Model/Catalog Number: 801402. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of t…
- Z-0242-2025 · Open, ClassifiedSURG PAT XRAY 1X1 Model/Catalog Number: 801403. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the…
- Z-0243-2025 · Open, ClassifiedSURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues …
- Z-0244-2025 · Open, ClassifiedSURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of t…
- Z-0245-2025 · Open, ClassifiedSURG PAT XRAY 1/2X3 Model/Catalog Number: 801407. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of t…
- Z-0246-2025 · Open, ClassifiedSURG PAT XRAY 1X3 Model/Catalog Number: 801408. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the…
- Z-0247-2025 · Open, ClassifiedSURG PAT XRAY 3X3 Model/Catalog Number: 801409. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the…
- Z-0248-2025 · Open, ClassifiedSURG STRP 1/8X6 Model/Catalog Number: 801449. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the c…
- Z-0249-2025 · Open, ClassifiedSURG STRP 1/4X6 Model/Catalog Number: 801450. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the c…
- Z-0250-2025 · Open, ClassifiedSURG STRP 1/2X6 Model/Catalog Number: 801451. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the c…
- Z-0251-2025 · Open, ClassifiedSURG STRP 3/4X6 Model/Catalog Number: 801452. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the c…
- Z-0252-2025 · Open, ClassifiedSURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the…
- Z-0253-2025 · Open, ClassifiedSURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the…
- Z-0255-2025 · Open, ClassifiedSURG STRP 3X6 Model/Catalog Number: 801456. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the cen…
- Z-0256-2025 · Open, ClassifiedSURG STRP 3-1/2X6 Model/Catalog Number:801457. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page