SenTiva, REF: 1000C, Generators Part of the VNS Therapy… recallZ-1073-2025 · LivaNova USA, Inc. · initiated Dec 12, 2024
Class II Open, Classified All recalls posted in January 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from LivaNova USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
FDA root cause category: Device Design
What the company is doing
On 12/12/2024, correction notices were mailed to customers asking them to do the following: 1. Monitor patient frequently and perform an on-demand System Diagnostic test at each visit per labeling (typical patient visit cadence is every 3-12 months). See VNS Therapy Physician's Manual, found in the Manuals Section https://www.livanova.com/epilepsy-vnstherapy/en-us/hcp/product-manuals or https://www.livanova.com/depression/en-us/healthcare-provider-resources. 2. Counsel prospective and current patients to ensure that they are aware of and understand the risks. You may elect to provide the patient letter. 3. Contact Customer Quality at 1-866-882-8804 or e-mail at [email protected] if communication with the generator using the Programming System indicates that: i. Diagnostic tests cannot be performed (Error Code 254 - Test Interrupted error message) ii. System Diagnostic test displays a LOW lead impedance result. 4. Encourage patients to: i. Notify you if there is a change in perceived clinical symptoms (e.g. increase in seizures or depressive symptoms, etc.). ii. Notify you if they no longer perceive any form of stimulation. 5. Ensure that this notice is communicated to all personnel within your organization who need to be aware. 6. Complete and return the response form via email to [email protected] Firm is updating the VNS Therapy Instructions for Use (IFU) to reflect the patient management recommendations relevant to this issue. For questions contact the firm's Customer Quality at 1-866-882-8804 or e-mail [email protected] On 6/16/2025, updated recall notices were delivered in person notifying customers that devices with a design change were available so affected generators remaining in inventory should be replaced. New devices with the design change have serial numbers greater than 500,000. There are no new actions for devices that have already been implanted.
Recall details
- Recall number
- Z-1073-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2024
- Posted by the FDA
- Jan 31, 2025
- Recalling firm
- LivaNova USA, Inc.
- Quantity
- 48
- Product code
- LYJ
- Distribution
- US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.
Product description
SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).
Code information (models, lots, serial numbers)
UDI-DI: 05425025750405. All Serial Numbers less than 500,000.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VNS Therapy® SenTiva™ Model 1000 GeneratorLivanova USA, Inc. · Stimulator, Autonomic Nerve, Implanted For Epilepsy3 recalls188 deaths3,378 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95435).
- Z-1072-2025 · Open, ClassifiedSenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page