The FLUOROspeed is intended to be used as a universal… recallZ-0317-2025 · Shimadzu Corporation Medical Systems Division · initiated Sep 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shimadzu Corporation Medical Systems Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

FDA root cause category: Software change control

What the company is doing

On 9/13/24, recall notices were emailed to consignees asking them to do the following: 1) If the problem occurs it is possible to solve this problem by turning off the power of the system and then pushing the reset switch on the Digital radiography unit s control cabinet. 2) Your firm authorized service representative will be contacting to schedule the implementation of the corrective action but if you are not contacted, contact the firm's National Support. 3) Complete and return the response form via email to response to [email protected] If you have questions contact the firm's National Technical Support at 844-487-2767, available 8:00AM-5:00PM, Monday - Friday.

Recall details

Recall number
Z-0317-2025
Status
Open, Classified
Initiated by the company
Sep 6, 2024
Posted by the FDA
Nov 7, 2024
Recalling firm
Shimadzu Corporation Medical Systems Division
Quantity
30
Product code
JAA
FDA submissions
K191877
Distribution
US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.

Product description

The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.

Code information (models, lots, serial numbers)

UDI-DI: 04540217065271. Serial Numbers: MQ493E3E3001, MQ493E3E3002, MQ493E3D6002, MQ493E3D5001, MQ493E3D8002, MQ493E3E1002, MQ493E3D4002, MQ493E3D6003, MQ493E3DA002, MQ493E3E2001, MQ493E3E2002, MQ493E3E1001, MQ493E3DC001, MQ493E3D9001, MQ493E3D4005, MQ493E3D7001, MQ493E3DB002, MQ493E3D2002, MQ493E3D4001, MQ493E3D7002, MQ493E3D4003, MQ493E3D4004, MQ493E3D2001, MQ493E3D8001, MQ493E3D5002, MQ493E3DB001, MQ493E3D2003, MQ493E3DA001, MQ493E3D6001, MQ493E3E1003

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95450).