GE Healthcare Giraffe Incubator Carestation CS1, Infant… recallZ-0141-2025 · Datex--Ohmeda, Inc. · initiated Jun 6, 2023
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datex--Ohmeda, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
FDA root cause category: Component design/selection
What the company is doing
GE HealthCare notified its consignees of the issue by letter beginning on 06/06/2023 using a traceable means. The notice explained the issue of potentially higher levels of formaldehyde which rapidly decrease of a week, and instructions on how to accelerate the rate of dissipation if the unit had not yet been put into clinical use. For questions: contact 1-800-437-1171 or your local GE HealthCare Representative
Recall details
- Recall number
- Z-0141-2025
- Status
- Completed
- Initiated by the company
- Jun 6, 2023
- Posted by the FDA
- Oct 31, 2024
- Recalling firm
- Datex--Ohmeda, Inc.
- Quantity
- 224 units
- Product code
- FMZ
- FDA submissions
- K213553, K152809
- Distribution
- worldwide
Product description
GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
1) 2082844-002-01040417, UDI: (01)00840682116855(21)TAAC70003(11)230106, (01)00840682116855(21)TAAC70004(11)230109, (01)00840682116855(21)TAAC70011(11)230110, (01)00840682116855(21)TAAC70012(11)230110, (01)00840682116855(21)TAAC70013(11)230111, (01)00840682116855(21)TAAC70015(11)230112, (01)00840682116855(21)TAAC70016(11)230112, (01)00840682116855(21)TAAC70020(11)230112, (01)00840682116855(21)TAAC70022(11)230112, (01)00840682116855(21)TAAC70024(11)230112, (01)00840682116855(21)TAAC70147(11)230317, (01)00840682116855(21)TAAC70148(11)230317, (01)00840682116855(21)TAAC70149(11)230317, (01)00840682116855(21)TAAC70150(11)230317, (01)00840682116855(21)TAAC70152(11)230320, (01)00840682116855(21)TAAC70153(11)230320, (01)00840682116855(21)TAAC70155(11)230322, (01)00840682116855(21)TAAC70158(11)230322, (01)00840682116855(21)TAAC70159(11)230323, (01)00840682116855(21)TAAC70160(11)230323, (01)00840682116855(21)TAAC70161(11)230323, (01)00840682116855(21)TAAC70162(11)230323, (01)00840682116855(21)TAAC70163(11)230324, (01)00840682116855(21)TAAC70164(11)230324, (01)00840682116855(21)TAAC70165(11)230324, (01)00840682116855(21)TAAC70166(11)230324, (01)00840682116855(21)TAAC70167(11)230327, (01)00840682116855(21)TAAC70169(11)230327, (01)00840682116855(21)TAAC70170(11)230327, (01)00840682116855(21)TAAC70171(11)230328, (01)00840682116855(21)TAAC70177(11)230330, (01)00840682116855(21)TAAC70178(11)230330, (01)00840682116855(21)TAAC70179(11)230331, (01)00840682116855(21)TAAC70181(11)230331, (01)00840682116855(21)TAAC70182(11)230331, (01)00840682116855(21)TAAC70184(11)230331, (01)00840682116855(21)TAAC70192(11)230405, (01)00840682116855(21)TAAC70193(11)230405, (01)00840682116855(21)TAAC70196(11)230405, (01)00840682116855(21)TAAC70197(11)230405, (01)00840682116855(21)TAAC70200(11)230407, (01)00840682116855(21)TAAC70201(11)230407, (01)00840682116855(21)TAAC70202(11)230407, (01)00840682116855(21)TAAC70203(11)230407, (01)00840682116855(21)TAAC70204(11)230407, (01)00840682116855(21)TAAC70205(11)230407, (01)00840682116855(21)TAAC70208(11)230411, (01)00840682116855(21)TAAC70209(11)230411, (01)00840682116855(21)TAAC70210(11)230411, (01)00840682116855(21)TAAC70211(11)230414, (01)00840682116855(21)TAAC70212(11)230414, (01)00840682116855(21)TAAC70213(11)230414, (01)00840682116855(21)TAAC70214(11)230414, (01)00840682116855(21)TAAC70215(11)230417, (01)00840682116855(21)TAAC70216(11)230417, (01)00840682116855(21)TAAC70217(11)230417, (01)00840682116855(21)TAAC70218(11)230417, (01)00840682116855(21)TAAC70223(11)230418, (01)00840682116855(21)TAAC70224(11)230418, (01)00840682116855(21)TAAC70225(11)230418, (01)00840682116855(21)TAAC70226(11)230418, (01)00840682116855(21)TAAC70227(11)230418, (01)00840682116855(21)TAAC70228(11)230418, (01)00840682116855(21)TAAC70229(11)230419, (01)00840682116855(21)TAAC70242(11)230421, (01)00840682116855(21)TAAC70243(11)230421, (01)00840682116855(21)TAAC70244(11)230421, (01)00840682116855(21)TAAC70245(11)230421, (01)00840682116855(21)TAAC70246(11)230421, (01)00840682116855(21)TAAC70247(11)230421; 2) 2082844-002-01044956, UDI: (01)008406821168552(21)TAA23140007SA(11)230404, (01)008406821168552(21)TAA23140008SA(11)230404; 3) 2082844-002-01058424, UDI: (01)00840682116855(21)TAAC70142(11)230213; 4) 2082844-002-01069017, UDI: (01)008406821168552(21)TAA23110001SA(11)230316, (01)008406821168552(21)TAA23110002SA(11)230316, (01)008406821168552(21)TAA23110003SA(11)230316; 5) 2082844-002-01089666, UDI: (01)00840682116855(21)TAAC70124(11)230201; 6) 2082844-002-01090474, UDI: (01)00840682116855(21)TAAC70126(11)230201, (01)00840682116855(21)TAAC70128(11)230206, (01)00840682116862(21)TAAC70132(11)230206, (01)00840682116855(21)TAAC70133(11)230206, (01)00840682116855(21)TAAC70125(11)230201, (01)00840682116855(21)TAAC70127(11)230206, (01)00840682116855(21)TAAC70129(11)230206, (01)00840682116855(21)TAAC70131(11)230206, (01)00840682116855(21)TAAC70136(11)230206, (01)00840682116855(21)TAAC70137(11)230206; 7) 2082844-002-01091269, UDI: (01)008406821168552(21)TAA23090001SA(11)230228 228; 8) 2082844-002-01093489, UDI: (01)00840682116855(21)TAAC70187(11)230403, (01)00840682116855(21)TAAC70188(11)230403, (01)00840682116855(21)TAAC70189(11)230403, (01)00840682116855(21)TAAC70190(11)230403, (01)00840682116855(21)TAAC70191(11)230403; 9) 2082844-002-01113262, UDI: (01)008406821168552(21)TAA23050010SA(11)230130, (01)008406821168552(21)TAA23050011SA(11)230130, (01)008406821168552(21)TAA23050012SA(11)230130, (01)008406821168552(21)TAA23050013SA(11)230130; 10) 2082844-002-01113711, UDI: (01)008406821168552(21)TAA23130005SA(11)230328, (01)008406821168552(21)TAA23130007SA(11)230329, (01)008406821168552(21)TAA23130008SA(11)230329, (01)008406821168552(21)TAA23130009SA(11)230329; 11) 2082844-002-01123085, UDI: (01)008406821168552(21)TAA23190005SA(11)230513, (01)008406821168552(21)TAA23190006SA(11)230513, (01)008406821168552(21)TAA23190007SA(11)230513, (01)008406821168552(21)TAA23190008SA(11)230513, (01)008406821168552(21)TAA23190009SA(11)230513, (01)008406821168552(21)TAA23190010SA(11)230513; 12) 2082844-002-01123918, UDI: (01)008406821168552(21)TAA23130001SA(11)230328; 13) 2082844-002-01128966, UDI: (01)008406821168552(21)TAA23040022SA(11)230127; 14) 2082844-002-01133042, UDI: (01)008406821168552(21)TAA23130003SA(11)230328, (01)008406821168552(21)TAA23130004SA(11)230328, (01)008406821168552(21)TAA23130006SA(11)230329; 15) 2082844-002-01146447, UDI: (01)008406821168552(21)TAA23100001SA(11)230307; 16) 2082844-002-100205, UDI: (01)008406821168552(21)TAA23030030SA(11)230121; 17) 2082844-002-100749, UDI: (01)00840682116855(21)TAAC70272(11)230426; 18) 2082844-002-100752, UDI: (01)00840682116855(21)TAAC70075(11)230123, (01)00840682116855(21)TAAC70083(11)230124, (01)00840682116855(21)TAAC70084(11)230124, (01)00840682116855(21)TAAC70086(11)230125, (01)00840682116855(21)TAAC70087(11)230125; 19) 2082844-002-100995, UDI: (01)008406821168552(21)TAA23090003SA(11)230228, (01)008406821168552(21)TAA23090004SA(11)230228, (01)008406821168552(21)TAA23090005SA(11)230228, (01)008406821168552(21)TAA23090006SA(11)230228, (01)008406821168552(21)TAA23090002SA(11)230228, (01)008406821168552(21)TAA23090007SA(11)230228; 20) 2082844-002-101425, UDI: (01)008406821168552(21)TAA23190003SA(11)230512, (01)008406821168552(21)TAA23160001SA(11)230419, (01)008406821168552(21)TAA23170001SA(11)230424, (01)008406821168552(21)TAA23130031SA(11)230331, (01)008406821168552(21)TAA23130032SA(11)230331, (01)008406821168552(21)TAA23130033SA(11)230331, (01)008406821168552(21)TAA23130034SA(11)230331, (01)008406821168552(21)TAA23130035SA(11)230331, (01)008406821168552(21)TAA23130036SA(11)230331, (01)008406821168552(21)TAA23130021SA(11)230330, (01)00840682116855(21)TAAC70143(11)230224, (01)00840682116855(21)TAAC70144(11)230224, (01)00840682116855(21)TAAC70145(11)230224, (01)008406821168552(21)TAA23130025SA(11)230331, (01)008406821168552(21)TAA23130026SA(11)230331, (01)008406821168552(21)TAA23130027SA(11)230331, (01)008406821168552(21)TAA23130028SA(11)230331, (01)008406821168552(21)TAA23130029SA(11)230331, (01)008406821168552(21)TAA23130030SA(11)230331, (01)00840682116855(21)TAAC70146(11)230313, (01)008406821168552(21)TAA23140001SA(11)230403, (01)008406821168552(21)TAA23140002SA(11)230403, (01)008406821168552(21)TAA23130022SA(11)230330, (01)008406821168552(21)TAA23130023SA(11)230330, (01)008406821168552(21)TAA23130024SA(11)230330, (01)008406821168552(21)TAA23160002SA(11)230419; 21) 2082844-002-523548, UDI: (01)00840682116855(21)TAAW62043(11)170130; (01)00840682116855(21)TAAZ70271(11)200224; 22) 2082844-002-529928, UDI: (01)008406821168552(21)TAA23130002SA(11)230328; 23) 2082844-002-594800, UDI: (01)00840682116855(21)TAAB70352(11)220520; 24) 2082844-002-643544, UDI: (01)008406821168552(21)TAA23030023SA(11)230120, (01)008406821168552(21)TAA23030024SA(11)230120, (01)008406821168552(21)TAA23030025SA(11)230120; 25) 2082844-002-925841, UDI: (01)00840682116855(21)TAAB70547(11)220815; 26) 2082844-002-965540, UDI: (01)00840682116855(21)TAAB70617(11)220826, (01)00840682116855(21)TAAB70665(11)220928, (01)00840682116855(21)TAAB70666(11)220928, (01)00840682116855(21)TAAB70670(11)220929, (01)00840682116855(21)TAAB70671(11)220929, (01)00840682116855(21)TAAB70681(11)221003, (01)00840682116855(21)TAAB70691(11)221004, (01)00840682116855(21)TAAB70702(11)221005, (01)00840682116855(21)TAAB70703(11)221005, (01)00840682116855(21)TAAB70732(11)221010; 27) 2082844-002-982916, UDI: (01)00840682116855(21)TAAC70033(11)230113, (01)00840682116855(21)TAAC70037(11)230116, (01)00840682116855(21)TAAC70044(11)230116, (01)00840682116855(21)TAAC70052(11)230119, (01)00840682116855(21)TAAC70058(11)230119, (01)00840682116855(21)TAAC70059(11)230119, (01)00840682116855(21)TAAC70060(11)230119, (01)00840682116855(21)TAAC70065(11)230120, (01)00840682116855(21)TAAC70067(11)230123, (01)00840682116855(21)TAAC70068(11)230123, (01)00840682116855(21)TAAB70608(11)220824; 28) 2082844-002-983420, UDI: (01)00840682116855(21)TAAC70076(11)230123, (01)00840682116855(21)TAAC70078(11)230124, (01)00840682116855(21)TAAC70079(11)230124, (01)00840682116855(21)TAAC70082(11)230124, (01)00840682116855(21)TAAC70089(11)230125, (01)00840682116855(21)TAAC70090(11)230125, (01)00840682116855(21)TAAC70092(11)230126, (01)00840682116855(21)TAAC70093(11)230126; 29) 2082844-002-993905, UDI: (01)008406821168552(21)TAA23040004SA(11)230123, (01)008406821168552(21)TAA23040005SA(11)230123, (01)008406821168552(21)TAA23040006SA(11)230124, (01)008406821168552(21)TAA23050014SA(11)230130; 30) 2082844-002-995018, UDI: (01)00840682116855(21)TAAB70551(11)220815, (01)00840682116855(21)TAAB70558(11)220819; 31) 2082844-002-995799, UDI: (01)008406821168552(21)TAA23040019SA(11)230127, (01)008406821168552(21)TAA23040020SA(11)230127, (01)008406821168552(21)TAA23190004SA(11)230513, (01)008406821168552(21)TAA23160007SA(11)230420, (01)008406821168552(21)TAA23160009SA(11)230420, (01)008406821168552(21)TAA23130015SA(11)230329, (01)008406821168552(21)TAA23130016SA(11)230329, (01)008406821168552(21)TAA23130017SA(11)230329, (01)008406821168552(21)TAA23130018SA(11)230329, (01)008406821168552(21)TAA23130020SA(11)230329, (01)008406821168552(21)TAA23160003SA(11)230419, (01)008406821168552(21)TAA23160004SA(11)230419, (01)008406821168552(21)TAA23160005SA(11)230419, (01)008406821168552(21)TAA23160006SA(11)230419, (01)008406821168552(21)TAA23160008SA(11)230420, (01)008406821168552(21)TAA23160010SA(11)230420, (01)008406821168552(21)TAA23160011SA(11)230420, (01)008406821168552(21)TAA23200003SA(11)230515, (01)008406821168552(21)TAA23200004SA(11)230515, (01)008406821168552(21)TAA23200005SA(11)230515, (01)008406821168552(21)TAA23200006SA(11)230515, (01)008406821168552(21)TAA23200007SA(11)230515, (01)008406821168552(21)TAA23200008SA(11)230515, (01)008406821168552(21)TAA23200009SA(11)230515, (01)008406821168552(21)TAA23200010SA(11)230515, (01)008406821168552(21)TAA23130019SA(11)230329; 32) 2082844-002-998943, UDI: (01)00840682116855(21)TAAC70050(11)230119, (01)00840682116855(21)TAAC70053(11)230119, (01)00840682116855(21)TAAC70110(11)230130
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Giraffe OmniBed Carestation CS1Datex-Ohmeda, Inc. · INCUBATOR, NEONATAL2 recalls217 reports
- Giraffe Incubator Carestation CS1Datex-Ohmeda, Inc. · INCUBATOR, NEONATAL1 recall37 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95456).
- Z-0140-2025 · CompletedGE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.