ClariTEE Probe- Miniature transesophageal echo ultrasound… recallZ-0231-2025 · ImaCor Inc. · initiated Sep 16, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ImaCor Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

FDA root cause category: Process control

What the company is doing

ImaCor issued Inventory Correction for ImaCor ClariTEE Probe (CLT-010 and CLT-010-1) letter on 9/16/24. Letter states reason for recall, health risk and action to take: 1. Identify the Affected Devices o Please check your inventory for ClariTEE probes with lot numbers 20250509 and 20250823. Inform ImaCor of the number of applicable probes currently in inventory. 2. Prepare the Products for Return o Package the affected products securely to prevent damage during shipment. 3. Shipping Instructions o Please ship the affected products via FedEx using FedEx Account Number 665380920 to cover the shipping cost. Ship to: ImaCor Inc, 50 Jericho Turnpike, Suite 105, Jericho, NY 11753. Destination phone number is (516) 393-0970. o Attach a copy of this letter inside the package for reference. 4. Replacement Devices o Once we receive the returned devices, we will ship replacement probes to your facility. Replacement probes are expected to be available on or about October 15, 2024. Please return all affected devices by October 7. If you have any questions or need assistance, please contact Jenn Kujawski at [email protected]. Our team is ready to assist you with any inquiries and can provide additional details if needed.

Recall details

Recall number
Z-0231-2025
Status
Open, Classified
Initiated by the company
Sep 16, 2024
Posted by the FDA
Oct 25, 2024
Recalling firm
ImaCor Inc.
Quantity
73 units
Product code
ITX
FDA submissions
K220490, K080223, K100989, K142054
Distribution
FL ,GA, NJ, TN

Product description

ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010, CLT-010-1

Code information (models, lots, serial numbers)

UDI - Primary - 00861589000108 Lot numbers: 20250509 and 20250823