Skip to content

iotaSOFT Plus Drive Unit, REF IM-05. The iotaSOFT Insertion… recallZ-0288-2025 · Iotamotion Inc · initiated Sep 12, 2024

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Iotamotion Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect GTIN number.

FDA root cause category: Process control

What the company is doing

The firm did not issue any communication. They distributed the devices and saw through their ERP system that the devices had been used. The remainder of the devices were in the possession of the firm's VP, therefore, not viewed as "distributed."

Recall details

Recall number
Z-0288-2025
Status
Completed
Initiated by the company
Sep 12, 2024
Posted by the FDA
Nov 1, 2024
Recalling firm
Iotamotion Inc
Quantity
5 devices
Product code
QQH
Distribution
Distribution was made to Iowa. There was no government/military/foreign distribution.

Product description

iotaSOFT Plus Drive Unit, REF IM-05. The iotaSOFT Insertion System is intended to aid the surgeon in placement of cochlear implant electrode arrays into a radiographically normal cochlea by controlling the speed of implant insertion.

Code information (models, lots, serial numbers)

Lot #D500330, EXP. 2025-01-12, UDI-DI 00850002960006 (the incorrect GTIN number listed on the label).

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page