Skip to content

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting… recallZ-0285-2025 · Zimmer, Inc. · initiated Sep 17, 2024

Class II Open, Classified All recalls posted in November 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

FDA root cause category: Component change control

What the company is doing

Zimmer Biomet, in partnership with Saphena Medical, Inc issued Urgent Medical Device Recall Letter- Lot Specific on 9/17/24 via Fed'X. Letter states reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, immediately locate and quarantine affected product in your inventory. 3. Return all affected product from your facility. a. Request a Return Goods Authorization (RGA) number by emailing the completed Attachment 1 Certificate of Acknowledgement form to [email protected]. b. After the RGA number is received, proceed with returning your inventory to the address below: PRIMO MEDICAL GROUP ATTN: ZB Saphena Medical Recall 75 Mill Street Stoughton, MA 02072 c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons 4. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 5. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. Upon receipt of affected product, Zimmer Biomet will credit your account. 6. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 7. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].

Recall details

Recall number
Z-0285-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 17, 2024
Posted by the FDA
Nov 1, 2024
Recalling firm
Zimmer, Inc.
Quantity
432 units
Product code
GEI
FDA submissions
K140603
Distribution
Nationwide

Product description

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Code information (models, lots, serial numbers)

UDI: +M718VPX40001/$$326051543DD6738Q, +M718VPX4001/$$326062143DD8092J Lot Numbers: 43DD6738, 43DD8092

Other devices under the same FDA submission

1 device cleared under K140603 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page