microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot… recallZ-0312-2025 · FHC, Inc. · initiated Oct 4, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from FHC, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

FDA root cause category: Process change control

What the company is doing

On October 4, 2024, the firm notified affected consignees via Urgent Medical Device Recall letters. Customers were instructed to immediately identify and quarantine all affected product. The firm will contact customers and issue an RMA to return affected product and issue a replacement by 10/11/24. Customers should not use the product in the interim. Should you have any further questions regarding this matter you may contact the firm by email at [email protected] or by telephone at 207-666-5425.

Recall details

Recall number
Z-0312-2025
Status
Open, Classified
Initiated by the company
Oct 4, 2024
Posted by the FDA
Nov 8, 2024
Recalling firm
FHC, Inc.
Quantity
8 packs of 5ea (40 total)
Product code
HAW
FDA submissions
K011775
Distribution
Domestic distribution only to the following states: AZ, CA, PA, TN, TX, WA.

Product description

microTargeting Insertion Tube Set, P/N 66-IT-05P, Lot 244517. The intended use is to accurately guide a microelectrode or instrument in the brain during stereotactic functional neurosurgical procedures.

Code information (models, lots, serial numbers)

Package DI: 10873263004323; Contains DI Package: 00873263004326 LOT No. 244517

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K011775 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page