Skip to content

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80… recallZ-0315-2025 · Biocare Medical, LLC · initiated Sep 25, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biocare Medical, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 9/25/24, recall notices were sent to customers and distributors who were asked to do the following: 1) Locate any affected kits in your possession and discontinue use. 2) Share this notification with all users within your facility and network to ensure all personnel who may use the kits are aware of this recall. 3) Discard any product remaining in your possession. 4) Complete and return the Response Form via email to [email protected] Distributors were asked to send the customer notice to their customers. Questions regarding this recall notification may be forwarded to [email protected] On 12/12/2024, recall notices were sent to customers containing additional affected lots.

Recall details

Recall number
Z-0315-2025
Status
Open, Classified
Initiated by the company
Sep 25, 2024
Posted by the FDA
Nov 6, 2024
Recalling firm
Biocare Medical, LLC
Quantity
976
Product code
PPM
Distribution
US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.

Product description

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20

Code information (models, lots, serial numbers)

REF/UDI-DI/Kit Lot(DAB Lot): IPK5011G80/00847627009005/042324(012424A),060523(051823A), 082923(062923A), 011624(072723A), 042324(012424A), 080524(032624A); OPRI6056KT180/00847627024985/020624(111623A),032224(111623A-3); IPK5010G80/00847627008992/071923(051823A), 091923(081023A), 112823(101923A), 100424(011524A), 041624(022624A); OPRI6056KT180/00847627024985/071924(061824A); IPI4006KG10/00847627008985/080323(062923A), 102723(081023A)

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates