SafeControl handset (a component of the Sentida bed which… recallZ-0329-2025 · Wissner-Bosserhoff GMBH · initiated Jul 11, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Wissner-Bosserhoff GMBH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.

FDA root cause category: Device Design

What the company is doing

Wissner-bosserhoff issued an Urgent safety information notice along with supplement use instructions notice to its sole US consignee on 07/11/2022. The notice explained the problem, risk to patient, and requested the user update the IFU for the device. Distributors were directed to notify their customers.

Recall details

Recall number
Z-0329-2025
Status
Open, Classified
Initiated by the company
Jul 11, 2022
Posted by the FDA
Nov 8, 2024
Recalling firm
Wissner-Bosserhoff GMBH
Quantity
77 units
Product code
LLI
Distribution
US Nationwide distribution in the states of NC.

Product description

SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198

Code information (models, lots, serial numbers)

UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156