Avoximeter Whole Blood Oximeter recallZ-0279-2025 · Accriva Diagnostics, Inc. · initiated Sep 27, 2024

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Accriva Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.

FDA root cause category: Process change control

What the company is doing

On 9/27/24, a recall notice was sent to a customer informing them to do the following: 1) This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where the potentially affected device has been transferred.. 2. Call Technical Support 1-800-678-0710 to initiate a Service Call and Returned Material Authorization 3. Complete and return the customer response form via email to For technical questions contact Technical Support at the number above or email [email protected].

Recall details

Recall number
Z-0279-2025
Status
Completed
Initiated by the company
Sep 27, 2024
Posted by the FDA
Oct 31, 2024
Recalling firm
Accriva Diagnostics, Inc.
Quantity
1
Product code
DQA
FDA submissions
K922075
Distribution
US: AZ

Product description

AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Code information (models, lots, serial numbers)

UDI-DI: 10711234530023. Serial Number: 6824

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K922075 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.