Skip to content

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis… recallZ-0286-2025 · Siemens Healthcare Diagnostics Inc · initiated Sep 24, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.

FDA root cause category: Under Investigation by firm

What the company is doing

An Urgent Medical Device Correction letter dated September 2024 was sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " If you are a distributor, please ensure your customers receive this UMDC letter. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " Adverse reactions or quality problems experienced with the use of the affected products listed in the above table may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " For any questions, you may call Siemens Healthineers Technical Solutions Center at 877-229-3711.

Recall details

Recall number
Z-0286-2025
Status
Open, Classified
Initiated by the company
Sep 24, 2024
Posted by the FDA
Nov 1, 2024
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
1292 units
Product code
CHL
FDA submissions
K200107
Distribution
US Nationwide. Canada.

Product description

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Code information (models, lots, serial numbers)

UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K200107 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.