Skip to content

DeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06 recallZ-0309-2025 · DeRoyal Industries Inc · initiated Oct 3, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DeRoyal Industries Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.

FDA root cause category: Under Investigation by firm

What the company is doing

DeRoyal issued an URGENT! DEROYAL RECALL NOTICE to its consignees on 10/03/2024 via email. The notice explained the issue, potential risk, and requested the return of the affected packs. If the affected product was further distributed, the notice directed the distributor to notify its customers. This may be achieved directly by forwarding the recall notification and notify DeRoyal that you have done so, or the distributor may provide DeRoyal with your customer listing allowing them to contact your customers.

Recall details

Recall number
Z-0309-2025
Status
Open, Classified
Initiated by the company
Oct 3, 2024
Posted by the FDA
Nov 5, 2024
Recalling firm
DeRoyal Industries Inc
Quantity
95 units
Product code
FDE
Distribution
US Nationwide distribution in the state of OH.

Product description

DeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06

Code information (models, lots, serial numbers)

UDI/ DI 00749756913393 (each), 50749756913398 (case): Lot Numbers: lot 59244907 exp 10/1/2025

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95509).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page