Baxter AK 98 Dialysis Machine, Hemodialysis Delivery… recallZ-0327-2025 · Baxter Healthcare Corporation · initiated Oct 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Baxter sent an IMPORTANT MEDICAL DEVICE CORRECTION notice to its consignees on 10/16/2024 via USPS. Until Baxter has transitioned to newly manufactured platinum-cured silicone tubing components in AK 98 machines, the firm recommends these actions: - If you have new AK 98 machines with a runtime less than 500 hours, consider using other available machines including AK 98 machines with a runtime greater than 500 hours, as PCB and PCBA levels decrease over time of use. - If no alternative is available, continue to use the AK 98 to ensure your patients have access to dialysis treatment. - Do not stop dialysis treatment to patients who need it. - Call Baxter Renal Customer Care at 800-525-2623, option 2, option 2, between 7:00 am and 5:00 pm CT, Monday thru Friday to provide the serial numbers of your AK 98 machines with a runtime less than 500 hours to determine a correction plan for your site. - Until the solution is implemented to reduce the possible presence of PCB and PCBA below the FDA recommended limits, Baxter is holding shipment of new AK 98 machines, allowing for the treatment needs of the dialysis population in our customers care. If you have a medical need for a new AK 98 machine with the peroxide-cured silicone tubing, contact your Baxter Renal Sales Representative for instructions on placing the order. - Please forward a copy of this communication to other facilities or departments that have this product. ***Updated 12/3/25*** On 12/3/2025, Vantive sent a follow-up Important Medical Device Correction letter. This letter contained an update to health hazard based on Vantive's toxicological risk assessment. The issue is unlikely to pose a risk for patients with a body weight greater than 25kg, irrespective of the run-time of the machine. The firm has determined that no corrections are needed for devices currently with customers, and patients may continue therapy with the current configuration of AK 98.
Recall details
- Recall number
- Z-0327-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 16, 2024
- Posted by the FDA
- Nov 13, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 1100 units
- Product code
- KDI
- FDA submissions
- K232467
- Distribution
- US Nationwide distribution and worldwide
Product description
Baxter AK 98 Dialysis Machine, Hemodialysis Delivery System, Product Code 955607. As of February 2025, the firm name is Vantive.
Code information (models, lots, serial numbers)
UDI/DI 07332414124519, All serials with runtime less than 500 hours
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AK 98 955607Vantive Health GmbH · Dialyzer, High Permeability With Or Without Sealed Dialysate System1 recall1 death6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page