Skip to content

Galaxy System (GAL-001) recallZ-0742-2025 · Noah Medical · initiated Oct 28, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Noah Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

FDA root cause category: Component change control

What the company is doing

On 10/28/24, recall notices were sent to customers who were told the firm would contact them to arrange inspection and replacement of the power supply system. Share this notification with any other healthcare professionals who may use the system. Complete and return the response form. Firm customer Service: [email protected], 888-325-6624

Recall details

Recall number
Z-0742-2025
Status
Open, Classified
Initiated by the company
Oct 28, 2024
Posted by the FDA
Dec 20, 2024
Recalling firm
Noah Medical
Quantity
18
Product code
EOQ
FDA submissions
K223144
Distribution
US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

Product description

Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

Code information (models, lots, serial numbers)

UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532,

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates