Defibtech DDP-200P: Pediatric Defibrillation Pads 0 For use… recallZ-0503-2025 · Defibtech, LLC · initiated Oct 14, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Defibtech, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
FDA root cause category: Under Investigation by firm
What the company is doing
Defibtech issued Urgent Field Safety Notice on 10/14/24 by e-mail to the Irish Market h via the Distributor Oxygen Care Ltd. titled: Unauthorized Label Applied to Defibtech, LLC's Automated External Defibrillator (AED} Pads within the Irish Market. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/User: 1. Check the label of your Defibtech AED Pads. Please review Attachment B which clearly shows authorised labelled Defibtech AED Pads versus unauthorised labelled Defibtech AED Pads. 2. Please confirm receipt of this notice by completing Attachment C "Customer Reply Form", attached to this notice, and return to Oxygen Care, Ltd (our new official Irish Distributor) so that our records may be updated. 3. If an unauthorised label applied to Defibtech AED Pads is confirmed, please urgently return the Defibtech AED Pads along with the completed "Customer Reply Form" to Oxygen Care. Upon receipt of the unauthorized labelled Defibtech AED Pads, Oxygen Care will send a replacement set of authorised labelled Defibtech AED Pads to you at no charge: Oxygen Care Ltd.: Email: Sales@ oxyge n-care.ie Phone: 01 2769700 Address: 2 Holfeld Business Park, Kilmacanogue. Co. Wicklow. Eircode A98 NP27 4. If you have transferred the Defibtech AED Pads to another entity or department, please contact Oxygen Care as per the above contact details. The unauthorised labelled Defibtech AED pads were re-labelled with expiration dates beyond the shelf life of the Defibtech AED Pads. The affected Defibtech AED Pads should not be used. Defibtech will continue to work closely with regulatory authorities toward the resolution of this issue. Defibtech is committed to patient safety and is focused on providing the highest level of customer satisfaction. We appreciate your cooperation and sincerely apologize for any inconvenience this may cause you.
Recall details
- Recall number
- Z-0503-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 14, 2024
- Posted by the FDA
- Nov 25, 2024
- Recalling firm
- Defibtech, LLC
- Quantity
- 22,294 total
- Product code
- MKJ
- FDA submissions
- P160032
- Distribution
- Ireland
Product description
Defibtech DDP-200P: Pediatric Defibrillation Pads 0 For use with AED Models: Lifeline and Lifeline Auto (DDU-l XX Series)-Indicated for use on victims of sudden cardiac arrest (SCA.)
Code information (models, lots, serial numbers)
UDI-DI: 00815098020430, DDP-200P pediatric defibrillation pads, single pair; 10815098020437, DDP-200P pediatric defibrillation pads, single pair configuration; 40815098020438, DDP-200P pediatric defibrillation pads, 20 pairs
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- DDU-1XX Pediatric Rescue Pads DDP-200PDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall0 reports
- DDU-1XX Series Pediatric Pads Configuration DDP-200PDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall0 reports
Other devices under the same FDA submission
2 devices cleared under P160032 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Lifeline VIEW AUTO Configuration DDU-E2200Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)2 recalls11 reports
- Lifeline VIEW AUTO DDU-E2200Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)2 recalls8 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95594).
- Z-0502-2025 · Open, ClassifiedDefibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XX Series)- Indicated for use on victims of sudden cardiac arrest (SCA)
- Z-0504-2025 · Open, ClassifiedDefibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victi…
- Z-0505-2025 · Open, ClassifiedDefibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page