Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F… recallZ-0506-2025 · Zyno Medical LLC · initiated Oct 18, 2024

Class II Open, Classified All recalls posted in November 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zyno Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

FDA root cause category: Device Design

What the company is doing

Zyno Medical notified consignees on 10/18/2024 via email. Consignees were instructed that the updated notification supersedes current labeling and includes the following instructions: To maintain the health and runtime of the battery, it is recommended to plug the battery into AC Mains in for at least 6 hours every 30 days to fully charge the battery. Additionally, for scenarios where the Z800 pump is left plugged into AC Mains for extended periods of time, it is recommended to unplug for up to 5 seconds and re-plug the Z800 pump back into AC Mains at least every 30 days to activate the internal battery charger and ensure an expected operating time on battery power. Customers were also asked to provide acknowledgement of receipt.

Recall details

Recall number
Z-0506-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Oct 18, 2024
Posted by the FDA
Nov 21, 2024
Recalling firm
Zyno Medical LLC
Quantity
34,994 units
Product code
FRN
FDA submissions
K130690, K100705
Distribution
This product is distributed only in the United States.

Product description

Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.

Code information (models, lots, serial numbers)

Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200.

Other devices under the same FDA submission

4 devices cleared under K130690, K100705 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.