Skip to content

Acrobat-I Stabilizer recallZ-0486-2025 · Maquet Cardiovascular, LLC · initiated Oct 16, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.

FDA root cause category: Process change control

What the company is doing

MAQUET Cardiovascular LLC / Getinge notified consignees via letter FedEx 2-Day Delivery on 10/16/2024. Customers were instructed to forward the notification to all users and notify customers if further distributed, examine inventory and return any affected inventory, and to complete and return the MEDICAL DEVICE REMOVAL - RESPONSE FORM.

Recall details

Recall number
Z-0486-2025
Status
Open, Classified
Initiated by the company
Oct 16, 2024
Posted by the FDA
Nov 18, 2024
Recalling firm
Maquet Cardiovascular, LLC
Quantity
33,823 units
Product code
MWS
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.

Product description

Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model Number C-OM-10000Z. UDI-DI Number: 00607567100008. Batch Numbers: 3000277221, 3000286549, 3000315868, 3000317517, 3000322538, 3000339447, 3000363028, 3000270713, 3000275435, 3000275771, 3000280868, 3000284390, 3000303149, 3000321642, 3000342528, 3000336070, 3000322179, 3000337396, 3000273182, 3000282181, 3000337880, 3000353535, 3000364172, 3000357109, 3000276109, 3000338107, 3000354445, 3000355050, 3000352899, 3000324785, 3000360894, 3000370720, 3000323627, 3000346330, 3000356740, 3000278665, 3000344012, 3000366729, 3000361801, 3000400928, 3000363775, 3000311956, 3000343553, 3000367559, 3000377818, 3000341430, 3000367345, 3000336377, 3000316208, 3000381449, 3000380956, 3000403756, 3000321414, 3000361499, 3000407809, 3000398030, 3000387678, 3000320077, 3000277072, 3000283754, 3000296552, 3000320660, 3000339096, 3000347694, 3000348390, 3000352250, 3000279303, 3000277309, 3000282794, 3000283482, 3000350599, 3000351774, 3000351996, 3000365741, 3000369223, 3000369738, 3000393258, 3000323129, 3000273547, 3000366133, 3000391146, 3000273843, 3000280827, 3000295912, 3000346410, 3000401855, 3000384033, 3000382967, 3000378703, 3000363755, 3000316146, 3000295949, 3000398861, 3000394893, 3000380542, 3000361797, 3000350598, 3000320073, 3000299042, 3000295186, 3000288696, 3000284723, 3000274876, 3000363227, 3000285664.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95614).