MEDLINE procedure kits labeled as follows recallZ-0513-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Oct 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The connectors may not allow catheters to be inserted as intended if the connector is not in the fully open position.
FDA root cause category: Under Investigation by firm
What the company is doing
An IMMEDIATE ACTION REQUIRED RECALL notification letter dated 10/16/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place unaffected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-216-FGRecall Code: 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0513-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 16, 2024
- Posted by the FDA
- Nov 22, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 3744 units
- Product code
- LRO
- Distribution
- US Nationwide. Bermuda, Canada, Panama, UAE, US Virgin Islands
Product description
MEDLINE procedure kits labeled as follows: 1) CONT EPI 20G W/HUSTEAD, REF PAIN1069B; 2) CSE TRAY, REF PAIN1941; 3) TRAY CON EPID 20G NYL W/TUOHY, REF PAIN9023; 4) TRAY, CONT EPI 20G NYL W/ HUST, REF PAIN1281 5) TRAY, CONT EPI 20G NYL W/ HUST, REF PAIN9024
Code information (models, lots, serial numbers)
Codes: 1) PAIN1069B UDI-DI: 10889942757381 (each) 40889942757382 (case), Lot Numbers: 22GLA433, 23CLA257; 2) PAIN1941 UDI-DI: 10195327365417 (each) 40195327365418(case), Lot Numbers: 23ELB081; 3) PAIN9023 UDI-DI: 10889942448715(each) 40889942448716 (case), Lot Numbers: 22GLA470, 22ILA287; 4) PAIN1281 UDI-DI: 10888277748408(each) 40888277748409(case), Lot Numbers: 22FLA647; 5) PAIN9024 UDI-DI: 10889942448708(each) 40889942448709(case), Lot Numbers: 22FLA675, 22GLA241, 22HLB225, 22ILA288
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. PAIN9024Medline Industries, Inc. · General surgery tray2 recalls2 reports
- Medline Industries, Inc. PAIN9023Medline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. PAIN1941Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. PAIN1281Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. PAIN1069BMedline Industries, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95615).
- Z-0508-2025 · Open, ClassifiedMEDLINE NEXUS 19/20G CATH CONNECTOR, Medline SKU REF NEXCONNECT, Catheter Connector
- Z-0509-2025 · Open, ClassifiedMEDLINE procedure kits labeled as follows: 1) COMBINED SPINAL AND EPIDURAL, REF DYNJRA1268B; 2) COMBINED SPINAL AND EPIDURAL, REF DYNJRA1268C; 3) FEMORAL BLOCK TRAY, REF DYNJRA…
- Z-0510-2025 · Open, ClassifiedMEDLINE procedure kits labeled as follows: 1) 20G EPI CATH SINGLE STERILE, REF EPICATH19 2) 20G EPI CATH SINGLE STERILE, REF EPICATH20
- Z-0511-2025 · Open, ClassifiedMEDLINE procedure kits labeled as follows: 1) ECHOSTIM BLOCK SET, REF DYNJRA9043S; 2) ULTRASOUND BLOCK PREP TRAY, REF PAIN1622A
- Z-0512-2025 · Open, ClassifiedMEDLINE procedure kits labeled as follows: 1) PERIPHERAL CONTINUOUS NERVE BL, REF PAIN1423A; 2) CONT. 20G EPI MRI COMPATIBLE, REF PAIN1470A; 3) EPIDURAL CATHETERIZATION KIT, REF P…
- Z-0514-2025 · Open, ClassifiedMEDLINE procedure kits labeled as follows: 1) NEURAXIAL KIT, REF PAIN1876A
- Z-0515-2025 · Open, ClassifiedMEDLINE procedure kits labeled as follows: 1) L&D CONTINUOUS EPIDURAL TRAY, REF PAIN1699; 2) L&D CONTINUOUS EPIDURAL TRAY, REF PAIN1699A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.