Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC… recallZ-0553-2025 · Shimadzu Corp. Analytical & Measuring Instruments Div. · initiated Oct 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Shimadzu Corp. Analytical & Measuring Instruments Div. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
FDA root cause category: Device Design
What the company is doing
On 10/9/24, field safety notices were mailed to customers asking them to do the following: (1) From the front and the right sides, check to see if there is any condensation water leakage. (2) If any condensation water leakage is found in the inspection, install Dew Tray Replace Kit (S228-61501-41). (3) Regardless of the checking result, attach the Main PCB Protective Sponge (S228-69971-41) For questions, please call 800-477-1227 between the hours of 8:30-5:00 EST or visit www.ssi.shimadzu.com.
Recall details
- Recall number
- Z-0553-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 9, 2024
- Posted by the FDA
- Nov 26, 2024
- Recalling firm
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Quantity
- 179
- Product code
- KIE
- Distribution
- Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.
Product description
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Model/UDI-DI/Serial Number: SIL-20ACXR CL/04540217420131/L20995250001, L20995250002, L20995250003, L20995250004, L20995250005, L20995250006, L20995250007, L20995250008, L20995350010, L20995350011, L20995350012, L20995350013, L20995350014, L20995350015, L20995450016, L20995450017, L20995450018, L20995450019, L20995450020, L20995450021, L20995450022, L20995450023, L20995450024, L20995450025, L20995450026, L20995550027, L20995550028, L20995550029, L20995550030, L20995650031, L20995650032, L20995650033, L20995650034, L20995650035, L20995650036, L20995650037, L20995650038, L20995650039, L20995650040, L20995650041, L20995650042, L20995650043, L20995650044, L20995650045, L20995650046, L20995560001, L20995660002, L20995660003, L20995660004, L20995660005, L20995660006, L20995660007, L20995660008, L20995660009, L20995660010, L20995660011, L20995660012, L20995660013, L20995660014, L20995660015, L20995660016, L20995660017, L20995660018, L20995660019, L20995660020, L20995660021, L20995660022, L20995660023, L20995660024, L20995660025, L20995660026, L20995660027, L20995660028, L20995660029, L20995660030, L20995660031, L20995660032, L20995660033, L20995660034, L20995660035, L20995660036, L20995660037, L20995660038, L20995660039, L20995660040, L20995660041, L20995660042, L20995660043, L20995660044, L20995660045, L20995660046, L20995660047, L20995660048, L20995660049, L20995660050, L20995660051, L20995660052, L20995660053, L20995660054, L20995660055, L20995660056, L20995660057, L20995250009, L20995270001, L20995270002, L20995270003, L20995270004, L20995270005, L20995370006, L20995370007, L20995370008, L20995370009, L20995370010, L20995370011, L20995370012, L20995370013, L20995370014, L20995370015, L20995370016, L20995370017, L20995370018, L20995370019, L20995370020, L20995370021, L20995370022, L20995370023, L20995370024, L20995370025, L20995470028, L20995770032. SIL-20ACHT CL/04540217420124/L20985350001, L20985450002, L20985750003, L20985750004. SIL-30AC CL/04540217420155/L21005250001, L21005250002, L21005250003, L21005250004, L21005250005, L21005250006, L21005250007, L21005350008, L21005350009, L21005350010, L21005350011, L21005350012, L21005350013, L21005350014, L21005350015, L21005350016, L21005350017, L21005350018, L21005350019, L21005350020, L21005350021, L21005350022, L21005350023, L21005350024, L21005450025, L21005550026, L21005550027, L21005650028, L21005650029, L21005750030, L21005750031, L21005570001, L21005570002, L21005670003, L21005670004, L21005670005, L21005670006, L21005670007, L21005670008, L21005770009, L21005770010, L21005770011, L21005770012, L21005770013, L21005770014.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SIL-20ACXR CL 228-45136-54Shimadzu Corporation · Apparatus, High Pressure Liquid Chromatography1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page