Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL recallZ-0554-2025 · Shimadzu Corp. Analytical & Measuring Instruments Div. · initiated Oct 9, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Shimadzu Corp. Analytical & Measuring Instruments Div. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

FDA root cause category: Device Design

What the company is doing

On 10/9/24, field safety notices were mailed to customers asking them to do the following: Check the Serial Numbers, and if the instruments are affected by this issue, contact your regional office. The corrective actions to be taken to affective products are (1) Add a cover part to the main board. (2) Update the SIL-40 firmware to Ver. 1.16 or above. For questions, please call 800-477-1227 between the hours of 8:30-5:00 EST or visit www.ssi.shimadzu.com.

Recall details

Recall number
Z-0554-2025
Status
Open, Classified
Initiated by the company
Oct 9, 2024
Posted by the FDA
Nov 26, 2024
Recalling firm
Shimadzu Corp. Analytical & Measuring Instruments Div.
Quantity
48
Product code
KIE
Distribution
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.

Product description

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model/UDI-DI/Serial Numbers: SIL-40C XR CL/04540217420582/L2C316051002, L2C316051003, L2C316051004, L2C316051005, L2C316051006, L2C316051007, L2C316151012, L2C316151013, L2C316151014, L2C316151015. SIL-40C X3 CL/04540217420575/L2C305850001, L2C305950002, L2C306151025, L2C306051001, L2C306051002, L2C306051003, L2C306051004, L2C306051005, L2C306051007, L2C306051006, L2C306051008, L2C306051009, L2C306051010, L2C306051011, L2C306051012, L2C306051013, L2C306051014, L2C306051015, L2C306151017, L2C306151018, L2C306151022, L2C306151023, L2C306151024, L2C306151026, L2C306151028, L2C306151027, L2C306151029, L2C306151030, L2C306151031, L2C306151032, L2C306151033, L2C306151034, L2C306151035, L2C306151036, L2C306151037, L2C306151038, L2C306151040, L2C306151039. Firmware versions below 1.16.