VITROS recallZ-0535-2025 · Ortho-Clinical Diagnostics, Inc. · initiated Oct 11, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.
FDA root cause category: Under Investigation by firm
What the company is doing
QuidelOrtho issued Urgent Product Correction Notification letter on 10.21.24. Letter states reason for recall, health risk and action to take: " Discontinue using, render unusable, and discard all remaining inventory of lots manufactured from the affected coating(s) of VITROS Chemistry Products CRP Slides. " Complete and return the enclosed Confirmation of Receipt form. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. " Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System in your laboratory until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Root Cause Our investigation is ongoing; however, additional testing has been implemented in our release process to mitigate this issue prior to the release of future coatings of VITROS CRP Slides
Recall details
- Recall number
- Z-0535-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 11, 2024
- Posted by the FDA
- Nov 25, 2024
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 7717 units
- Product code
- DCK
- FDA submissions
- K953197, K152433
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
Product description
VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740
Code information (models, lots, serial numbers)
UDI/GTIN 10758750000296 Coating 1221, expiration date 10/01/2024 Coating 1222, expiration date 01/01/2025
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 1926740Ortho-Clinical Diagnostics, Inc. · C-REACTIVE PROTEIN, ANTIGEN, ANTISERUM, AND CONTROL1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.