Atellica CI Analyzer. Catalog Numbers: 10947347 recallZ-0597-2025 · Siemens Healthcare Diagnostics, Inc. · initiated Nov 4, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

FDA root cause category: Software design

Recall details

Recall number
Z-0597-2025
Status
Open, Classified
Initiated by the company
Nov 4, 2024
Posted by the FDA
Dec 4, 2024
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
559 units
Product code
JJE
FDA submissions
K222116
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.

Product description

Atellica CI Analyzer. Catalog Numbers: 10947347.

Code information (models, lots, serial numbers)

UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K222116 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.