Skip to content

Invivo recallZ-0539-2025 · Philips North America Llc · initiated Nov 4, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential safety issue where a patient may be harmed while preparing for or during a scan.

FDA root cause category: Use error

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION notice dated 10/31/24 was mailed to consignees. Consignees are to continue using affected systems in accordance with the intended use. When using the system follow section PRECAUTIONS, CAUTIONS & WARNINGS in the Instructions For Use (IFU). Consignees are to circulate the recall notice to all users for their awareness of the potential issue. The Recall Response Form is to be provided back to Philips MR within 30 days at [email protected]. Customers with any questions are to contact their local Philips Representative.

Recall details

Recall number
Z-0539-2025
Status
Open, Classified
Initiated by the company
Nov 4, 2024
Posted by the FDA
Nov 25, 2024
Recalling firm
Philips North America Llc
Quantity
5,231 units
Product code
MOS
FDA submissions
K032576
Distribution
US Nationwide distribution.

Product description

dS Breast 7ch 1.5T

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

REF: 453530280881, 453530280882, 459801290431, 459801290432; UDI: 00884838066847, 00884838086111; Serial No. 507, 648, 409, 133, 596, 421, 613, 68, 121, 132, 757, 787, 622, 538, 687, 139, 357, 806, 838, 555, 615, 754, 813, 525, 70, 1755, 518, 594, 628, 662, 502, 653, 575, 629, 473, 281, 335, 367, 364, 567, 1672, 1683, 740, 144, 1673, 561, 1655, 445, 435, 1725, 1661, 823, 1720, 1736, 1711, 1788, 1774, 427, 584, 589, 616, 627, 618, 540, 472, 436, 583, 715, 812, 1713, 510, 1692, 400, 1681, 741, 119, 382, 1780, 641, 236, 184, 467, 830, 794, 712, 433, 450, 486, 520, 624, 574, 605, 755, 751, 827, 1700, 661, 714, 1706, 1740, 825, 190, 438, 29, 60, 72, 83, 128, 129, 198, 138, 154, 185, 191, 146, 283, 285, 362, 365, 300, 372, 323, 377, 666, 678, 535, 609, 683, 802, 781, 676, 1685, 795, 539, 610, 654, 1763, 1787, 790, 665, 681, 75, 122, 797, 1773, 422, 474, 453, 528, 550, 1739, 1734, 1718, 807, 837, 710, 1670, 734, 462, 482, 509, 579, 696, 348, 657, 829, 828, 836, 761, 818, 530, 634, 1674, 1735, 461, 526, 519, 562, 586, 593, 577, 645, 675, 658, 643, 670, 796, 744, 801, 1719, 800, 835, 1684, 1689, 1737, 1764, 1772, 735, 803, 1715, 388, 21, 90, 9, 2, 155, 109, 179, 197, 160, 172, 116, 169, 183, 194, 175, 201, 199, 286, 356, 391, 386, 387, 1771, 705, 1785, 1669, 1747, 778, 699, 644, 739, 644, 1680, 580, 651, 582, 604, 637, 603, 647, 667, 693, 732, 704, 749, 833, 792, 811, 1701, 429, 419, 438, 426, 453, 468, 487, 504, 524, 553, 573, 542, 548, 557, 558, 614, 576, 632, 625, 650, 652, 548, 728, 669, 698, 730, 700, 729, 711, 743, 799, 760, 1675, 1656, 1646, 1693, 1724, 1722, 1729, 1775, 747, 89, 95, 100, 106, 76, 20, 207, 114, 117, 159, 123, 200, 196, 209, 217, 212, 324, 343, 389, 358, 346, 378, 366, 381, 370, 399, 425, 394, 432, 430, 499, 551, 475, 663, 104, 148, 177, 158, 413, 411, 415, 455, 512, 1679, 566, 817, 779, 477, 789, 822, 1696, 152, 731, 67, 140, 349, 373, 690, 282, 464, 606, 585, 630, 816, 1697, 659, 396, 1, 30, 120, 166, 345, 375, 443, 488, 213, 597, 607, 638, 1782, 279, 815, 1647, 213, 590, 529, 1746, 655, 1784, 371, 390, 798, 777, 1648, 1691, 758, 1676, 1704, 736, 300, 533, 559, 588, 587, 118, 124, 186, 393, 416, 1702, 1742, 476, 523, 672, 1678, 1659, 1786, 824, 1738, 599, 454, 694, 1768, 668, 392, 398, 825, 1654, 819, 805, 1790, 531, 831, 1756, 460, 503, 554, 522, 552, 706, 791, 804, 1745, 1645, 161, 126, 137, 145, 189, 147, 410, 423, 545, 1690, 497, 689, 691, 1789, 458, 463, 569, 826, 1698, 620, 631, 125, 379, 572, 673, 1660, 397, 505, 1721, 1726, 110, 111, 176, 546, 543, 549, 578, 713, 437, 1705, 642, 1781, 446, 448, 660, 1770, 544, 418, 633, 188, 534, 440, 621, 613, 471, 501, 511, 679, 742, 688, 748, 793, 702, 820, 1712, 1748, 58, 112, 287, 374, 412, 780, 1714, 424, 466, 479, 485, 563, 556, 568, 677, 595, 611, 640, 656, 664, 646, 686, 703, 707, 750, 763, 709, 1677, 1682, 1738, 810, 762, 814, 692, 738, 759, 1703, 59, 484, 34, 42, 10, 88, 71, 18, 64, 43, 78, 103, 82, 98, 136, 152, 130, 135, 150, 119, 182, 151, 157, 187, 184, 180, 202, 192, 280, 153, 168, 241, 326, 208, 380, 327, 383, 368, 403, 384, 376, 385, 434, 420, 451, 470, 498, 532, 527, 536, 617, 764, 788, 1727, 1728, 1744, 113, 195, 363, 135, 565, 506, 591, 173, 165, 171, 3083, 3019, 3127, 3025, 3034, 3035, 3124, 3080, 3015, 3037, 3067, 3003, 3088, 3095, 3058, 3129, 3096, 3078, 3020, 3117, 3132, 3069, 3110, 3128, 3113, 3087, 3062, 3001, 3086, 3033, 3051, 3097, 3056, 3094, 825, 3054, 3125, 3106, 3116, 3036, 3052, 3030, 3005, 3105, 3131, 3104, 3090, 3137, 3134, 3118, 3021, 3091, 3041, 3061, 3068, 3046, 3018, 3009, 3010, 3042, 3055, 3060, 3011, 3040, 3065, 3047, 3053, 3098, 3130, 3139, 3112, 3108, 3063, 3059, 3032, 3007, 3103, 3061, 3089, 3023, 701, 3006, 3133, 1738, 3081, 3115, 3026, 3012, 3102, 3031, 3121, 3093, 3119, 3109, 3085, 3089, 3028, 3082, 3084, 3013, 3008, 3079, 3114, 3060, 3005, 3016, 3041, 3122, 3004, 3029, 3107, 3002, 3027, 195, 158, 357, 178, 334, 324, 85, 142, 318, 286, 149, 157, 128, 124, 176, 148, 195, 168, 96, 153, 131, 135, 141, 127, 163, 292, 134, 194, 193, 350, 254, 197, 81, 238, 291, 323, 199, 155, 267, 338, 317, 314, 354, 366, 378, 340, 25, 363, 121, 188, 336, 145, 360, 20, 3, 122, 183, 196, 234, 316, 329, 89, 70, 353, 118, 233, 231, 288, 247, 369, 21, 133, 130, 241, 356, 68, 119, 351, 320, 380, 190, 299, 69, 78, 94, 72, 285, 295, 297, 283, 365, 330, 160, 181, 266, 335, 185, 174, 298, 287, 319, 184, 284, 364, 250, 303, 200, 82, 227, 119, 165, 201, 328, 157, 229, 242, 248, 251, 156, 294, 246, 144, 231, 245, 186, 120, 99, 74, 90, 88, 123, 92, 151, 152, 162, 125, 166, 182, 345, 243, 249, 255, 265, 252, 342, 325, 331, 341, 300, 327, 307, 346, 377, 170, 304, 229, 348, 371, 236, 337, 259, 343, 306, 187, 164, 173, 22, 177, 310, 154, 370, 237, 313, 258, 83, 91, 305, 308, 355, 347, 321, 362, 257, 359, 67, 93, 349, 244, 253, 98, 235, 26, 2, 232, 1, 80, 75, 95, 66, 116, 169, 368, 161, 63, 230, 115, 64, 309, 147, 73, 114, 264, 150, 289, 296, 302, 367, 361, 379, 228, 261, 191, 315, 167, 143, 71, 86, 84, 179, 301, 293, 248, 326, 263, 311, 358, 352, 62, 146, 239, 262, 23, 79, 171, 256, 344, 312, 339, 1140, 1086, 1166, 1177, 1182, 1245, 1265, 1191, 1347, 1159, 1244, 1199, 1173, 1292, 1203, 1100, 1168, 1238, 1198, 1091, 1142, 1106, 1104, 1141, 1068, 1065, 1146, 1092, 1136, 1248, 1241, 1257, 1165, 1178, 1192, 1235, 1284, 1247, 1108, 1103, 1169, 1193, 1287, 1196, 1134, 1119, 1132, 1204, 1174, 1063, 1109, 1101, 1164, 1202, 1260, 1114, 1139, 1288, 1291, 1285, 1125, 1070, 1074, 1144, 1282, 1345, 1277, 1148, 1286, 1062, 1100, 1135, 1232, 1206, 1087, 1110, 1113, 1171, 1069, 1145, 1077, 1068, 1089, 1167, 1079, 1180, 1201, 1067, 1161, 1121, 1184, 1163, 1188, 1197, 1263, 1194, 1239, 1266, 1200, 1256, 1099, 1105, 1186, 1156, 1259, 1080, 1240, 1098, 1123, 1162, 1283, 1137, 1085, 1294, 1276, 1154, 1064, 1152, 1147, 1149, 1160, 1258, 1102, 1236, 1275, 1111, 1084, 1195, 1075, 1073, 1150, 1128, 1078, 1187, 1257, 1252, 1254, 1151, 1081, 1129, 1269, 1279, 1097, 1122, 1290, 1234, 1293, 1117, 1205, 1183, 1107, 1126, 1072, 1096, 1112, 1088, 1251, 1243, 1158, 1066, 1118, 1175, 1130, 1242, 1289, 1197, 1270, 1071, 1185, 1264, 1273, 1131, 1120, 1082, 1138, 1233, 1093, 1190, 1181, 1133, 1076, 1281, 1124, 1271, 1237, 1267, 1116, 1094, 1176, 1262, 1249, 1115, 1095, 1143, 1189, 1090, 1155.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95645).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page