BD BBL Sensi Disc Ampicillin 2 g (AM-2) - In-Vitro BD BBL… recallZ-0621-2025 · Becton Dickinson & Co. · initiated Nov 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
FDA root cause category: Under Investigation by firm
What the company is doing
BD issued URGENT: Medical Device Product Correction (IDS-24-5142) letter on 11/6/24 to Distributors (included Customer Letter) via e-mail, Letter states reason for recall, health risk and action to take: Actions for Clinical Users: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood by those within your organization. 3. There are no additional recommendations for repeat testing or review of prior results. 4. Share and post this customer letter with users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 5.Complete the attached Customer Response Form and return to your distributor, whether or not you have any of the impacted material. If you require further assistance, please contact: BD Contact Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted Mon Fri 8:00am and 5:00pm CT or Email: [email protected] Product Complaints, Technical Questions Post-Market Quality [email protected] Recall Questions
Recall details
- Recall number
- Z-0621-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2024
- Posted by the FDA
- Dec 5, 2024
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 13,190 units
- Product code
- JTN
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.
Product description
BD BBL Sensi Disc Ampicillin 2 g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
Code information (models, lots, serial numbers)
UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD BBL Sensi-Disc Ampicillin 2 ug 231263Becton, Dickinson And Company · SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page