Skip to content

BD BBL Sensi Disc Ampicillin 2 g (AM-2) - In-Vitro BD BBL… recallZ-0621-2025 · Becton Dickinson & Co. · initiated Nov 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

FDA root cause category: Under Investigation by firm

What the company is doing

BD issued URGENT: Medical Device Product Correction (IDS-24-5142) letter on 11/6/24 to Distributors (included Customer Letter) via e-mail, Letter states reason for recall, health risk and action to take: Actions for Clinical Users: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood by those within your organization. 3. There are no additional recommendations for repeat testing or review of prior results. 4. Share and post this customer letter with users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 5.Complete the attached Customer Response Form and return to your distributor, whether or not you have any of the impacted material. If you require further assistance, please contact: BD Contact Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted Mon Fri 8:00am and 5:00pm CT or Email: [email protected] Product Complaints, Technical Questions Post-Market Quality [email protected] Recall Questions

Recall details

Recall number
Z-0621-2025
Status
Open, Classified
Initiated by the company
Nov 6, 2024
Posted by the FDA
Dec 5, 2024
Recalling firm
Becton Dickinson & Co.
Quantity
13,190 units
Product code
JTN
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.

Product description

BD BBL Sensi Disc Ampicillin 2 g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263

Code information (models, lots, serial numbers)

UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page