ChemoLock Close Vial Spike REF CL-80 recallZ-0652-2025 · ICU Medical, Inc. · initiated Nov 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to finished goods incorrectly packaged and labeled.

FDA root cause category: Packaging process control

What the company is doing

On 11/06/2024, the firm send an "URGENT Medical Device Correction" Letter via third party (Sedgwick, Inc.) to customers informing them that product was incorrectly packaged and labeled. Product CL-80 package incorrectly contained product code CL-80S. The CL-80 is a used with a 20 mm or a 28 mm vial where as CL-80S is only use for a 20 mm vial. The issue is using the incorrect adapter may result in leakages. Actions to be taken by the Customer: Discontinue use mislabeled lot 13953938 of CL-80 and discard remaining product from this specific lot. 1. Check all inventory locations within your institution for the affected product listed in the notification and discontinue use. Quarantine and destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there./// 3. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification./ / 4. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. For questions/assistance, contact: Global Complaint Management 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or [email protected] To report adverse events or product complaints Customer Service 1-866-829-9025, option 8 or [email protected] (M-F, 8:00 am 6:00 pm CT) Additional information or to request a credit. Field Action Processing 1-888-843-0245 [email protected] Questions about this communication

Recall details

Recall number
Z-0652-2025
Status
Open, Classified
Initiated by the company
Nov 6, 2024
Posted by the FDA
Dec 9, 2024
Recalling firm
ICU Medical, Inc.
Quantity
6000 units
Product code
ONB
FDA submissions
K131549
Distribution
Worldwide - U.S. Nationwide distribution in the state of IL and the country of Canada.

Product description

ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The Chemolock is needle free and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.

Code information (models, lots, serial numbers)

Model/Catalog Number: CL-80 UDI code: (01)10887709057125(17)290401(30)50(10)13959938 Lot Number: 13959938

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

90 devices cleared under K131549 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page