Merlin.net Patient Care Network (PCN) Heart Failure Web… recallZ-0987-2025 · St. Jude Medical · initiated Nov 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from St. Jude Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
FDA root cause category: Under Investigation by firm
What the company is doing
On 11/06/2024, the firm sent an "URGENT: MEDICAL DEVICES CORRECTION", to customers to inform them that on September 30, 2024, the Merlin.net PCN HF Portal (MN6000) was migrated to the cloud database and following the migration, some customers reported that their Hospital Systems were unable to transmit to the Merlin.net PCN. User Action Requested: Abbott requests that clinicians take the following actions: The Impacted CM3100 Hospital System devices at your facility is/are listed in the enclosed Appendix. o For a Hospital System marked SESSIONS SENT no further correction is needed. o Hospital Systems marked UNSENT SESSIONS should be turned on to send unsent sessions: If your Hospital System has connectivity, it will immediately attempt to send after boot-up. A spinning icon will appear to indicate that transmission is in progress. If you see a button on the home screen showing X Unsent Sessions select the button and follow the prompts to send. If you are unable to send sessions, contact Abbott Remote Care Technical Support (RCTS) for assistance. Distribute this notice to appropriate personnel within your institution. Complete the included acknowledgment form and return it to Abbott. NOTE: Hospital Systems that were not used between September 30, 2024, and October 15, 2024, are not impacted. any questions about this notice or potential patient impact, please contact your Abbott representative or Remote Care Technical Support at 1-844-MYCMEMS (692-6367) Monday Friday, 5AM PST 5PM PST.
Recall details
- Recall number
- Z-0987-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2024
- Posted by the FDA
- Jan 17, 2025
- Recalling firm
- St. Jude Medical
- Quantity
- 25 systems
- Product code
- MOM
- FDA submissions
- P100045
- Distribution
- U.S.: CA, CT, GA, IL, MI, MN, NC, NJ, NY, OR, PA, TX, and WA. O.U.S.: N/A
Product description
Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart failure treatment.
Code information (models, lots, serial numbers)
Merlin.net Patient Care Network Heart Failure Web Application Model Number: MN6000 UDI-DI code: 05414734510035 CardioMEMS Hospital system Model: CM3100 Serial Numbers: M220200018 M220200028 M220300070 M220300173 M220300186 M220700135 M220900036 M220900045 M220900050 M220900052 M220900089 M220900109 M221100017 M221100020 M221100043 M221100060 M230500037 M230500038 M230900026 M230900037 M231100022 M240100031 M240100086 M240500086 M240500088
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Merlin.net MN6000Cardiomems, Inc. · Implantable pulse generator, pacemaker (non-CRT)4 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page