Galaxy System (GAL-001) recallZ-0693-2025 · Noah Medical · initiated Oct 25, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Noah Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
FDA root cause category: Device Design
What the company is doing
On 10/25/24, the firm started hand delivering correction notices to customers informing them of the following: If your device is affected the firm will contact you to inspect and if necessary to replace the power supply. Firm contact: [email protected], 1-888-325-6624
Recall details
- Recall number
- Z-0693-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 25, 2024
- Posted by the FDA
- Dec 18, 2024
- Recalling firm
- Noah Medical
- Quantity
- 26
- Product code
- EOQ
- FDA submissions
- K223144
- Distribution
- US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
Product description
Galaxy System (GAL-001)
Code information (models, lots, serial numbers)
UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB-1540, CRB-1541, CRB-1543, CRB-1554
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Galaxy System GAL-001Noah Medical Corporation · Bronchoscope (Flexible Or Rigid)5 recalls1 death35 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.