Skip to content

HotDog Temperature Management System recallZ-0700-2025 · Augustine Temperature Management, LLC · initiated Dec 4, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Augustine Temperature Management, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".

FDA root cause category: Device Design

What the company is doing

Augustine Temperature Management issued a Field Notification to its consignees on 12/04/2024 via email. The notice explained the issue, potential risk, and advised that there is an update to the User and Technical Manual to further clarify existing instructions for the device. If the device was further distributed, those distributing the device were directed to notify those to whom the device was distributed. Consignees are expected to respond via email. An optional power board update is available. For questions or concerns, contact: [email protected] or by telephone, 952-465-3500, fax, 952-465-3501, 8AM - 5PM Central time.

Recall details

Recall number
Z-0700-2025
Status
Open, Classified
Initiated by the company
Dec 4, 2024
Posted by the FDA
Dec 18, 2024
Recalling firm
Augustine Temperature Management, LLC
Quantity
135 units
Product code
DWJ
FDA submissions
K201779, K220941
Distribution
US and Canada

Product description

Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC71, 2) Model # WC71R (refurbished); used in conjunction with Warming Blankets and Mattresses.

Code information (models, lots, serial numbers)

1) UDI-DI 00850011479018, Serial Numbers: 20080004-23060141; 2) UDI-DI 00850011479285, Serial Numbers: 20080004-23060141

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K201779, K220941 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95703).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page