Skip to content

Hologic Sizer Set- sterilization tray and 6 individual… recallZ-0669-2025 · Hologic, Inc · initiated Oct 10, 2022

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hologic, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set

FDA root cause category: Labeling Change Control

What the company is doing

Hologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X. The letter states reason for recall, health risk and action to take: Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized. link to their Instructions for Use (IFU) below: https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf Please also note that when sterilizing F0101, one (1) layer of wrapping material (or one layer of 1 ply wrap) is necessary to provide a sterile barrier for each sterilization tray. Please do not utilize 2 layers or one layer of 2-ply wrap. In the future, Hologic will be updating the IFU supplied with the sizers within F0101, to reflect this change. mailto:[email protected]

Recall details

Recall number
Z-0669-2025
Status
Completed
Initiated by the company
Oct 10, 2022
Posted by the FDA
Dec 11, 2024
Recalling firm
Hologic, Inc
Quantity
178 units
Product code
FZX
Distribution
US Nationwide distribution.

Product description

Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101

Code information (models, lots, serial numbers)

Model No F0101. UDI 15420045515390. Lot Numbers: 186728, SS1-200727, SS1-181121, SS1-190301, SS1-190501, 175517, 175516, 162427, 140718-01, 171056, 175515.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates