Skip to content

StatStrip Glucose Hospital Meter System. Model/Catalog… recallZ-0681-2025 · Nova Biomedical Corporation · initiated Nov 20, 2024

Class I Open, Classified All recalls posted in December 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nova Biomedical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

FDA root cause category: Software design

What the company is doing

Nova Biomedical notified consignees via letter on 11/20/2024. Consignees were instructed that all current Nova StatStrip customers that have gone Live with their new meters recently received a meter software update (v.0.0.13.45 or above) from Nova Biomedical to eliminate the potential risk described above. All remaining customers that have been shipped StatStrip meters impacted by this field correction, will have their meter software updated by Nova Biomedical or local dealer prior to going "Live" with their new meters.

Recall details

Recall number
Z-0681-2025
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Nov 20, 2024
Posted by the FDA
Dec 20, 2024
Recalling firm
Nova Biomedical Corporation
Quantity
2,365 units
Product code
PZI
FDA submissions
K232075
Distribution
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.

Product description

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.

Code information (models, lots, serial numbers)

Model No 63685. UDI-DI (01)10385480636858. Software Versions: v0.0.13.10 to v0.0.13.44

Other devices under the same FDA submission

1 device cleared under K232075 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page