DRG:HYBRiD-XL Insulin. Model/Catalog Number: HYE-5361 recallZ-0921-2025 · DRG International, Inc. · initiated Oct 7, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DRG International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The user of the assay may either face a failure message during re-calibration of the Hybrid.XL instrument using the re-calibration R1 and R2 of the device or find controls out of specification (failed Control High) which causes a delay in results.

FDA root cause category: Nonconforming Material/Component

What the company is doing

DRG International began notifying consignees on 10/07/2024 via letter. Consignees were instructed to check inventory for affected units, cease use of them, and quarantine all product. Arrange return of those units. Consignees were also instructed to notify customers if units were further distributed.

Recall details

Recall number
Z-0921-2025
Status
Open, Classified
Initiated by the company
Oct 7, 2024
Posted by the FDA
Jan 15, 2025
Recalling firm
DRG International, Inc.
Quantity
10 units
Product code
CFP
Distribution
Worldwide - US Nationwide distribution in the state of New Jersey and the countries of Czechia and Romania.

Product description

DRG:HYBRiD-XL Insulin. Model/Catalog Number: HYE-5361

Code information (models, lots, serial numbers)

UDI: (01)04048474053616(17)20251130(10)429K084-1. Lot Number: 429K084-1