Medline brand, medical procedure convenience kits, labeled… recallZ-0726-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Nov 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
FDA root cause category: Nonconforming Material/Component
What the company is doing
An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-209-FGX1 Recall Code: Recall Code 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-0726-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2024
- Posted by the FDA
- Dec 20, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 22,970 kits
- Product code
- OJH
- Distribution
- US Nationwide. Canada.
Product description
Medline brand, medical procedure convenience kits, labeled as: 1) LAMINECTOMY CDS-LF, SKU CDS780045N; 2) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119U; 3) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119V; 4) CRANIOTOMY CDS, SKU CDS981753K; 5) LAMINECTOMY CDS-LF, SKU CDS981923X; 6) CRANIOTOMY, SKU CDS982719Q; 7) CRANIOTOMY, SKU CDS982719R; 8) CRANIOTOMY, SKU CDS982719S; 9) CRANIOTOMY, SKU CDS982719T; 10) CRANIOTOMY, SKU CDS982719U; 11) CRANIOTOMY, SKU CDS982719V; 12) CRANIOTOMY, SKU CDS982719W; 13) CRANIOTOMY, SKU CDS984074I; 14) SPINE, SKU CDS984943I; 15) SPINE, SKU CDS984943J; 16) KIT NEURO CRANIOTOMY FOR TUMOR, SKU DYKMBNDL144; 17) KIT NEURO CRANIOTOMY FOR TUMOR, SKU DYKMBNDL144A; 18) CRANIOTOMY #IMF 56646-LF, SKU DYNJVB91001; 19) NEURO SPINE #IMF 56655-LF, SKU DYNJVB91004; 20) CRANIOTOMY HARPER PACK-LF, SKU DYNJ0162022G; 21) CRANIOTOMY PACK-LF, SKU DYNJ0530906AP; 22) CRANIOTOMY PACK-LF, SKU DYNJ0530906AQ; 23) LAMINECTOMY PACK-LF, SKU DYNJ17452R; 24) LAMINECTOMY PACK-LF, SKU DYNJ17452S; 25) CRANIOTOMY PACK-LF, SKU DYNJ17455J; 26) CRANIOTOMY PACK-LF, SKU DYNJ17455K; 27) DEPAUL CRANIOTOMY-LF, SKU DYNJ20018P; 28) SPINAL PROCEDURE SC, SKU DYNJ33402K; 29) LAMINECTOMY PACK SC, SKU DYNJ35328J; 30) CRANIOTOMY PACK 9128175-LF, SKU DYNJ35982J; 31) LAMINECTOMY PACK, SKU DYNJ37683J; 32) SPINAL RECONSTRUCTION PACK, SKU DYNJ37689G; 33) CRANIOTOMY PACK, SKU DYNJ37714G; 34) CRANIOTOMY PACK, SKU DYNJ38317G; 35) SPINE PACK-LF, SKU DYNJ38889I; 36) ORTHO SPINE, SKU DYNJ40221A; 37) ORTHO SPINE, SKU DYNJ40221B; 38) ORTHO SPINE, SKU DYNJ40221C; 39) ORTHO SPINE, SKU DYNJ40221D; 40) ORTHO SPINE, SKU DYNJ40221F; 41) CRANIOTOMY PK, SKU DYNJ40388B; 42) CRANIOTOMY-LF, SKU DYNJ43319F; 43) NEURO SPINE PACK 319728, SKU DYNJ44628D; 44) CRANI PACK-LF, SKU DYNJ44805L; 45) LAMINECTOMY PACK-LF, SKU DYNJ44815J; 46) NEURO SPINE PACK, SKU DYNJ45419D; 47) NEURO SPINE PACK, SKU DYNJ45419F; 48) SPINE PACK, SKU DYNJ46317F; 49) CRANI PACK, SKU DYNJ47605B; 50) SCRIPPS SW CRANIOTOMY PACK-LF, SKU DYNJ49653C; 51) PK) NEURO-CRANI, SKU DYNJ53453A; 52) ACDF PACK-LF, SKU DYNJ57136A; 53) PEDI CRANI MAJOR, SKU DYNJ59338C; 54) UMC CRANIOTOMY, SKU DYNJ60497A; 55) UMC CRANIOTOMY PK, SKU DYNJ60497B; 56) CMC NEURO PACK-LF, SKU DYNJ61079C; 57) CMC NEURO PACK-LF, SKU DYNJ61079D; 58) CRANIOTOMY PK, SKU DYNJ65934B; 59) METREX PACK-LF, SKU DYNJ66104A; 60) CRANIOTOMY PACK, SKU DYNJ66419; 61) CRANIOTOMY PACK, SKU DYNJ66419A; 62) CRANIOTOMY PACK RFID, SKU DYNJ66419B; 63) CRANIOTOMY PACK, SKU DYNJ67355A; 64) CRANIOTOMY PACK, SKU DYNJ67355B; 65) OSC LAMINECTOMY PACK, SKU DYNJ67395; 66) SPINE PACK, SKU DYNJ67505F; 67) SPINE PACK, SKU DYNJ67505G; 68) SPINE PACK, SKU DYNJ67652A; 69) NEURO CRANI PACK, SKU DYNJ69752; 70) SPINE PACK, SKU DYNJ69753; 71) SPINE PACK, SKU DYNJ80550A; 72) CRANI PACK, SKU DYNJ80554A; 73) SPINE PACK, SKU DYNJ82105; 74) LAMINECTOMY I-LF, SKU DYNJ83783; 75) ANTERIOR CERVICAL-LF, SKU DYNJ83785; 76) NEURO-CRANI PK-LF, SKU DYNJ85475; 77) CRANIOTOMY PACK, SKU DYNJ85527A; 78) CRANIOTOMY-LF, SKU DYNJ900718D; 79) WMC ANTERIOR CERVICAL-LF, SKU DYNJ902443D; 80) WMC LAMINECTOMY-LF, SKU DYNJ902468J; 81) WMC LAMINECTOMY-LF, SKU DYNJ902468K; 82) WMC LAMINECTOMY-LF, SKU DYNJ902468L; 83) WMC LAMINECTOMY-LF, SKU DYNJ902468M; 84) WMC CRANIOTOMY ADD ON-LF, SKU DYNJ902473L; 85) CRANIOTOMY, SKU DYNJ903224R; 86) CRANIOTOMY, SKU DYNJ903224S; 87) CRANIOTOMY, SKU DYNJ903224T; 88) CRANIOTOMY, SKU DYNJ903224V; 89) DAVIS LAMI, SKU DYNJ903452D; 90) ORTHO INS-SPINE BACK, SKU DYNJ904661D; 91) ORTHO INS-SPINE ACDF, SKU DYNJ904662D; 92) CCOC SPINE EHRLER, SKU DYNJ904681F; 93) KIT) SPINE LH, SKU DYNJ906448B; 94) CRANI, SKU DYNJ906469F; 95) SPINE, SKU DYNJ907030; 96) JVL - MODULE BURR HOLE - CRAN, SKU DYNJ907147A; 97) BGMC MICRODISK-SPINAL, SKU DYNJ907349B; 98) KIT PEDI CRANI LAMI, SKU DYNJ907372A; 99) KIT SPINAL INSTRUMENTATION, SKU DYNJ907376; 100) SPINE, SKU DYNJ907819A; 101) UMC CRANIOTOMY, SKU DYNJ907945; 102)
Code information (models, lots, serial numbers)
Show the full code information (14k characters)
UDI-DI: CDS780045N, 10193489355406 (each), 40193489355407 (case), Lot Numbers: 22JBI816 22JBI817; UDI-DI: CDS780119U, 10193489371116 (each), 40193489371117 (case), Lot Numbers: 22CBL018 21KBV760 21KBV761 21IBV579 21HBB379 21CBC828 21ABJ206 20JBX601 20HBC289; UDI-DI: CDS780119V, 10195327114916 (each), 40195327114917 (case), Lot Numbers: 22FBU189 22DBN874; UDI-DI: CDS981753K, 10193489848946 (each), 40193489848947 (case), Lot Numbers: 22DLA844; UDI-DI: CDS981923X, 10195327434298 (each), 40195327434299 (case), Lot Numbers: 23HBS882; UDI-DI: CDS982719Q, 10193489634853 (each), 40193489634854 (case), Lot Numbers: 20LBD073 20KBE409 20IBD650 20IBE719 20GBT868 20EBI282; UDI-DI: CDS982719R, 10193489441031 (each), 40193489441032 (case), Lot Numbers: 22CBD762 22CBD763 21LBG182 21IBU365 21DBU895 21CBA765 21BBN538; UDI-DI: CDS982719S, 10195327148348 (each), 40195327148349 (case), Lot Numbers: 22GBX154 22GBX157 22FBN679; UDI-DI: CDS982719T, 10195327264161 (each), 40195327264162 (case), Lot Numbers: 23CBO275 23BBG742 23ABP314 23ABQ860 22KBQ362 22JBR890; UDI-DI: CDS982719U, 10195327359676 (each), 40195327359677 (case), Lot Numbers: 23GBO912 23FBP663; UDI-DI: CDS982719V, 10195327510831 (each), 40195327510832 (case), Lot Numbers: 24BBI145 24BBC547 24ABB119 23LBN381 23KBP461; UDI-DI: CDS982719W, 10195327645090 (each), 40195327645091 (case), Lot Numbers: 24DBS813; UDI-DI: CDS984074I, 10193489539240 (each), 40193489539241 (case), Lot Numbers: 22FBW145 22FBW156 22EBN997 22CBP097 21LBX349 21KBB576 21HBG033 21HBG034 21EBO565 21EBK739 21CBV476 20KBI591 20IBR095; UDI-DI: CDS984943I, 10193489805789 (each), 40193489805780 (case), Lot Numbers: 21DBG136 21BBN662 20LBC741 20KBB644 20HBX447 20GBU951; UDI-DI: CDS984943J, 10193489939828 (each), 40193489939829 (case), Lot Numbers: 22IBT623 22EBN175 22EBM178 22ELA400 22OBM639 22BBY198 22BBQ629 21VBB560 21IBD939 21FBR368; UDI-DI: DYKMBNDL144, 10193489478358 (each), 40193489478359 (case), Lot Numbers: 22HLA742 22FLA327 22ELB110 22DLA896 22DLA727 22CLA771 22BLA947 21LLA651 21LLA431 21KLB158 21JLA461 21ILA240 21HLA660 21GLA871 21GLA640 21FLA052 21DLA534 21DLA128 21CLA448 21CLA295 21CLA414 21CBH443 21BLA414 21ABU821 21ABI784 20LLA543; UDI-DI: DYKMBNDL144A, 10195327241032 (each), 40195327241033 (case), Lot Numbers: 23GLA491 23FLA815 23DLA832 23BLA654 22LLA124 22JLB035 22ILA386; UDI-DI: DYNJVB91001, 10193489801170 (each), 40193489801171 (case), Lot Numbers: 23HLA649 23GLA935 23GLA081 23CLA457 23BLA516 22ILB117 22ILA306 21HLA380 21HLA223; UDI-DI: DYNJVB91004, 10193489801163 (each), 40193489801164 (case), Lot Numbers: 23KLA445 23ELB201 23ELA545 23DLA441 23DLA190 23CLB159 23CLA458 23CLA132 23BLA369 23BLA190 23BLA230 22LLA793 22LLA465 22KLA564 22KLA338 22JLA726 22ILB118 22HLA309 22FLA968 22DLA954 22CLA724 22CLA439 22ALA872 22ALA757 21LLA754 21KLB076 21LLA003 21GLA530 21ELA378 21DLA724 21DLA250 21CLA168 21BLA553; UDI-DI: DYNJ0162022G, 10195327404598 (each), 40195327404599 (case), Lot Numbers: 23HDA239; UDI-DI: DYNJ0530906AP, 10195327629205 (each), 40195327629206 (case), Lot Numbers: 24DBU441 24CBD922 24BBG480; UDI-DI: DYNJ0530906AQ, 10198459023583 (each), 40198459023584 (case), Lot Numbers: 24HBF736 24GBF740; UDI-DI: DYNJ17452R, 10193489232745 (each), 40193489232746 (case), Lot Numbers: 20HBU778 20DBV450; UDI-DI: DYNJ17452S, 10193489448726 (each), 40193489448727 (case), Lot Numbers: 24DBD546 24BBJ956 24ABI951 23LBD415 23JBS986 23IBN443 23EBS949 23CBV576 23BBD547 23ABE395 22IBQ636 22IBG316 22HBD921 22FBN962 22EBF435 22OBF142 22ABJ199 21KBR795 21KBR798 21HBL602 21FBC898 21FBC897 21CBS933 21BBF488 21BBB399 20KBK827 20KBJ917; UDI-DI: DYNJ17455J, 10193489776034 (each), 40193489776035 (case), Lot Numbers: 20HBU443 20FBE933 20DBS073; UDI-DI: DYNJ17455K, 10193489448733 (each), 40193489448734 (case), Lot Numbers: 23FBR133 23CBH092 23CBE623 22LBN401 22FBK959 22NBF989 22BBK396 21IBK856 21BBA707 20LBS071; UDI-DI: DYNJ20018P, 10193489614862 (each), 40193489614863 (case), Lot Numbers: 22IBQ147 22HBX305; UDI-DI: DYNJ33402K, 10889942620814 (each), 40889942620815 (case), Lot Numbers: 22IBU396 22HBQ128; UDI-DI: DYNJ35328J, 10193489777376 (each), 40193489777377 (case), Lot Numbers: 22IBC437; UDI-DI: DYNJ35982J, 10195327109936 (each), 40195327109937 (case), Lot Numbers: 24DDB143 24CDC118 24CDB079 23KDA456 23IDC198 23GDA317 23ADA106 22LDB766 22GDA257 22GDA211 22DDB393; UDI-DI: DYNJ37683J, 10193489884579 (each), 40193489884570 (case), Lot Numbers: 22EMA087; UDI-DI: DYNJ37689G, 10193489888409 (each), 40193489888400 (case), Lot Numbers: 22FMA398; UDI-DI: DYNJ37714G, 10193489884586 (each), 40193489884587 (case), Lot Numbers: 23AMA818 22EMH802; UDI-DI: DYNJ38317G, 10195327188337 (each), 40195327188338 (case), Lot Numbers:22IMA064; UDI-DI: DYNJ38889I, 10193489433784 (each), 40193489433785 (case), Lot Numbers: 22FMC983; UDI-DI: DYNJ40221A, 10193489370706 (each), 40193489370707 (case), Lot Numbers: 20LMF699 20IMD325; UDI-DI: DYNJ40221B, 10193489468441 (each), 40193489468442 (case), Lot Numbers: 22DMF929 22BMD063 21KMA783 21EMB370; UDI-DI: DYNJ40221C, 10195327171094 (each), 40195327171095 (case), Lot Numbers: 22IMI082 22FMI349; UDI-DI: DYNJ40221D, 10195327281267 (each), 40195327281268 (case), Lot Numbers: 23FMA091 23BMH798 23BMD590 23AMH384 22KMB544; UDI-DI: DYNJ40221F, 10195327456931 (each), 40195327456932 (case), Lot Numbers: 24HMG815 24FMJ976 24FMB924 24EMF243 24DMH392 24CMF885 24AMC568 23LMD920 23KMH840 23JMI665 23HMF284 23GMG445; UDI-DI: DYNJ40388B, 10889942814510 (each), 40889942814511 (case), Lot Numbers: 22CBY753; UDI-DI: DYNJ43319F, 10193489379976 (each), 40193489379977 (case), Lot Numbers: 23GBU044 22HBJ618 22FBX102; UDI-DI: DYNJ44628D, 10195327336752 (each), 40195327336753 (case), Lot Numbers: 23GBR938; UDI-DI: DYNJ44805L, 10195327323516 (each), 40195327323517 (case), Lot Numbers: 23GBD055; UDI-DI: DYNJ44815J, 10195327323301 (each), 40195327323302 (case), Lot Numbers: 23GBM522 23GBE280 23EBI826 23CBI837; UDI-DI: DYNJ45419D, 10195327136116 (each), 40195327136117 (case), Lot Numbers: 23HMF531 23GMA066 23FMH176 23EMH480 23EMC171 23DMH420 23CMF514 23BMC506 22LMI321 22IMF730 22IMF731 22EMH912; UDI-DI: DYNJ45419F, 10195327625689 (each), 40195327625680 (case), Lot Numbers: 24HMI873 24DMK493; UDI-DI: DYNJ46317F, 10193489933154 (each), 40193489933155 (case), Lot Numbers: 22GBM895; UDI-DI: DYNJ47605B, 10193489730708 (each), 40193489730709 (case), Lot Numbers: 22IBF896; UDI-DI: DYNJ49653C, 10195327176730 (each), 40195327176731 (case), Lot Numbers: 22HMA233; UDI-DI: DYNJ53453A, 10193489227284 (each), 40193489227285 (case), Lot Numbers: 23GBW468; UDI-DI: DYNJ57136A, 10195327092160 (each), 40195327092161 (case), Lot Numbers: 22EMF805; UDI-DI: DYNJ59338C, 10195327021238 (each), 40195327021239 (case), Lot Numbers: 23HLA933; UDI-DI: DYNJ60497A, 10193489491494 (each), 40193489491495 (case), Lot Numbers: 21EMD822 20LMF845 20LME508; UDI-DI: DYNJ60497B, 10193489947649 (each), 40193489947640 (case), Lot Numbers: 23CBJ364 22GBQ933 22FBK324 22EBB521 22DBP934 21IBU612 21IBJ071; UDI-DI: DYNJ61079C, 10195327111335 (each), 40195327111336 (case), Lot Numbers: 23FBA577 23EBF442 23CBC033 23BBB993 22LBP597 22LBN337 22HBJ629 22GBD922 22EBC676; UDI-DI: DYNJ61079D, 10195327451561 (each), 40195327451562 (case), Lot Numbers: 23GBS127; UDI-DI: DYNJ65934B, 10193489918694 (each), 40193489918695 (case), Lot Numbers: 22DBF055; UDI-DI: DYNJ66104A, 10195327169480 (each), 40195327169481 (case), Lot Numbers: 22FMI384; UDI-DI: DYNJ66419, 10193489355468 (each), 40193489355469 (case), Lot Numbers: 20KBH494; UDI-DI: DYNJ66419A, 10195327105839 (each), 40195327105830 (case), Lot Numbers: 23IBO363 23DBF776 23CBU127 23BBM754 22LBN481 22HBP355 22HBP708 22GBW915 22EBV485 22DBQ127; UDI-DI: DYNJ66419B, 10195327670863 (each), 40195327670864 (case), Lot Numbers: 24GBC523 24EBT820; UDI-DI: DYNJ67355A, 10195327195441 (each), 40195327195442 (case), Lot Numbers: 23FLA443 23ELA388 23CLA826 23BLA666 23ALA391 22ILB109 22HLB181 22HLA296; UDI-DI: DYNJ67355B, 10195327456733 (each), 40195327456734 (case), Lot Numbers: 23ILA135 23HLA122 23GLA964; UDI-DI: DYNJ67395, 10193489895582 (each), 40193489895583 (case), Lot Numbers: 22FMD453 22FMB381; UDI-DI: DYNJ67505F, 10195327631734 (each), 40195327631735 (case), Lot Numbers: 24CBE725; UDI-DI: DYNJ67505G, 10195327670108 (each), 40195327670109 (case), Lot Numbers: 24HBM784 24GBD561 24EBS299 24CBT451; UDI-DI: DYNJ67652A, 10195327274542 (each), 40195327274543 (case), Lot Numbers: 23HME139; UDI-DI: DYNJ69752, 10193489977158 (each), 40193489977159 (case), Lot Numbers: 22EBJ914; UDI-DI: DYNJ69753, 10193489977233 (each), 40193489977234 (case), Lot Numbers: 22EBV579; UDI-DI: DYNJ80550A, 10195327064686 (each), 40195327064687 (case), Lot Numbers: 22FBW070; UDI-DI: DYNJ80554A, 10195327063313 (each), 40195327063314 (case), Lot Numbers: 22GBC812; UDI-DI: DYNJ82105, 10195327173074 (each), 40195327173075 (case), Lot Numbers: 23HBV574; UDI-DI: DYNJ83783, 10195327305918 (each), 40195327305919 (case), Lot Numbers: 23CBP256; UDI-DI: DYNJ83785, 10195327305932 (each), 40195327305933 (case), Lot Numbers: 23CBP258; UDI-DI: DYNJ85475, 10195327505905 (each), 40195327505906 (case), Lot Numbers: 24DMK076 23KMJ165; UDI-DI: DYNJ85527A, 10195327565176 (each), 40195327565177 (case), Lot Numbers: 23LBM088; UDI-DI: DYNJ900718D, 10195327686383 (each), 40195327686384 (case), Lot Numbers: 24HLA805 24GLB147 24GLA248; UDI-DI: DYNJ902443D, 10195327261276 (each), 40195327261277 (case), Lot Numbers: 24HBF620 24DBG611 23JBP842 23IBW095 23GBK736 23CBO765 23BBD677; UDI-DI: DYNJ902468J, 10195327261337 (each), 40195327261338 (case), Lot Numbers: 23ABB960 22JBY742; UDI-DI: DYNJ902468K, 10195327323462 (each), 40195327323463 (case), Lot Numbers: 23BBG726; UDI-DI: DYNJ902468L, 10195327372293 (each), 40195327372294 (case), Lot Numbers: 23IBO118 23HBK165 23FBO857 23DBN202; UDI-DI: DYNJ902468M, 10198459028151 (each), 40198459028152 (case), Lot Numbers: 24IBF034 24GBX534 24GBK426; UDI-DI: DYNJ902473L, 10195327261436 (each), 40195327261437 (case), Lot Numbers: 24ILA785 24ILA873 24ILA207 24HLA861 24HLA234 24GLA863 24GLA393 24ELA488 24DLA998 24CLA160 24ALA895 23LLA737 23KLB004 23KLA717 23KLA364 23ILA370 23GLA828 23GLA226 23FLA906 23DLA438 23CLA765 22LLA076; UDI-DI: DYNJ903224R, 10195327254261 (each), 40195327254262 (case), Lot Numbers: 22KDB985; UDI-DI: DYNJ903224S, 10195327328702 (each), 40195327328703 (case), Lot Numbers: 23IDA634 23BDB731 23ADC091; UDI-DI: DYNJ903224T, 10195327491086 (each), 40195327491087 (case), Lot Numbers: 24FDA232 24CDC017 24BDB603 23LDB727 23JDC541 23JDA679 23IDC174; UDI-DI: DYNJ903224V, 10198459060700 (each), 40198459060701 (case), Lot Numbers: 24IDA272; UDI-DI: DYNJ903452D, 10193489667820 (each), 40193489667821 (case), Lot Numbers: 22IBQ075; UDI-DI: DYNJ904661D, 10193489263190 (each), 40193489263191 (case), Lot Numbers: 23KBX356 23JBX848 23IBL881 23IBM306 23GBU556 23FBV382 23CBG855 23ABU277 22LBF701 22JBY022 22HBS591 22FBN765 22FBN766 22EBV206 22OBG856 22OBB277 22OBB280 21LBQ921 21IBU503 21GBC720 21FBF890 21DBU712 21CBE922 20EBT013; UDI-DI: DYNJ904662D, 10193489263206 (each), 40193489263207 (case), Lot Numbers: 23GBI899 23GBB212 23EBI632 23CBW838 23BBN983 23ABM490 22LBL126 22KBK688 22HBI868 22FBR270 22FBD458 22CLA847 22OBG841 22OBC072 22CLA235 21LBQ894 21JBM726 21GBS644 21FBJ147 21CBF257 21ABC418 20LBP879 20EBS185 20DBT259; UDI-DI: DYNJ904681F, 10193489803815 (each), 40193489803816 (case), Lot Numbers: 21EBC584 21BBB752 20KBL927 20GBC237 20FBC077; UDI-DI: DYNJ906448B, 10193489902921 (each), 40193489902922 (case), Lot Numbers: 22EBD157 22EBC254 22OBA537 22ABY321 21LBV221 21KBC470; UDI-DI: DYNJ906469F, 10195327210939 (each), 40195327210930 (case), Lot Numbers: 24HDA679 24FDA986 24BDB183 24BDB024 24ADA127 23IDC208 22KDA497 22JDB698 22IDB052 22HDC339; UDI-DI: DYNJ907030, 10193489325157 (each), 40193489325158 (case), Lot Numbers: 23BBU288 23ABT337 23ABS755 23ABN657 22JBB240 22EBI159 22EBG643 22EBH392 22DBE448 22OBI458 21LBU018 21LBU024 21EBO764 20JBO764 20JBL133 20DBO898; UDI-DI: DYNJ907147A, 10195327130770 (each), 40195327130771 (case), Lot Numbers: 22JBU269 22HBQ420 22GBR824; UDI-DI: DYNJ907349B, 10195327449568 (each), 40195327449569 (case), Lot Numbers: 23IBC888; UDI-DI: DYNJ907372A, 10195327060411 (each), 40195327060412 (case), Lot Numbers: 23JLA329; UDI-DI: DYNJ907376, 10193489889741 (each), 40193489889742 (case), Lot Numbers: 22GMB505; UDI-DI: DYNJ907819A, 10195327401054 (each), 40195327401055 (case), Lot Numbers: 23IBA532 23IBA146 23GBU799; UDI-DI: DYNJ907945, 10193489947762 (each), 40193489947763 (case), Lot Numbers: 23DBK646 23DBC953 22IBF953 22IBE829 22FBX834 22FBX835 22EBP537 22EBL544 22BBH770 21IBV572; UDI-DI: DYNJ908144, 10193489982190 (each), 40193489982191 (case), Lot Numbers: 21IBK042; UDI-DI: DYNJ908313, 10195327031459 (each), 40195327031450 (case), Lot Numbers: 23FBI071 23EBJ976 23ABP542 23ABE843 22LBB598 22JBT238 22JBA956 22EBQ482 22EBP641 21LBE373; UDI-DI: DYNJ908313A, 10195327436957 (each), 40195327436958 (case), Lot Numbers: 23IBT847 23HBT388 23GBO771; UDI-DI: DYNJ908699, 10195327159382 (each), 40195327159383 (case), Lot Numbers: 22HBO460; UDI-DI: DYNJ908964, 10195327225780 (each), 40195327225781 (case), Lot Numbers: 23FBC228 22KBD892; UDI-DI: DYNJ908964A, 10195327525811 (each), 40195327525812 (case), Lot Numbers: 24DBI341 24CBD458; UDI-DI: DYNJ909063A, 10195327460549 (each), 40195327460540 (case), Lot Numbers: 23HLB085; UDI-DI: DYNJ909364, 10195327319090 (each), 40195327319091 (case), Lot Numbers: 24CMC543 24BME891 24BME394 23LMF488 23LMB263 23IMH228 23IMD363 23HME707 23HME026 23CMI093
Devices named in this recall
108 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- Medline Industries, Inc. DYNJ17452SMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. CDS780119VMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ902473LMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJVB91001Medline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ59338CMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ907372AMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ67652AMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ0162022GMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. CDS982719VMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. CDS982719WMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ40388BMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ43319FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ38889IMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS982719RMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYKMBNDL144Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS984074IMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ20018PMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ903452DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ47605BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ35328JMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJVB91004Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ904681FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ37683JMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ37714GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ37689GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ907376Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ65934BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ46317FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS984943JMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ69752Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ80550AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ66419AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ35982JMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ61079CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ907147AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS982719SMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ40221CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ49653CMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67355AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ906469FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYKMBNDL144AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS982719TMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ909364Medline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS982719UMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ902468LMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. CDS981923XMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ908313AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ907349BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67355BMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ40221FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ45419FMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67505GMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ900718DMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ0530906AQMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ33402KMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ53453AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS780045NMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ66419Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40221AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ17455KMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40221BMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS982719QMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS981753KMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ67395Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906448BMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69753Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908144Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908313Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ80554AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ57136AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ66104AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ82105Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ38317GMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908964Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903224RMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902443DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902468JMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40221DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ83783Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ83785Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902468KMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903224SMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44628DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ907819AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61079DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ909063AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903224TMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ85475Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908964AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ85527AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0530906APMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ67505FMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ66419BMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902468MMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903224VMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ17452RMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ904661DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ904662DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ907030Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. CDS780119UMedline Industries, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95806).
- Z-0717-2025 · Open, ClassifiedMAJOR NEURO PACK-LF Convenience kit, SKU DYNJ0578916X. Convenience kits used for various procedures.
- Z-0718-2025 · Open, ClassifiedFESS-MB PACK-LF Convenience kit, SKU DYNJ0888159I. Convenience kits used for various procedures..
- Z-0719-2025 · Open, ClassifiedDIEP FLAP Convenience kit, SKU DYNJ910120. Convenience kits used for various procedures.
- Z-0720-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: 1) SINUS CDS-1, SKU CDS982901F; 2) NASAL PACK, SKU DYNJ17493G; 3) FESS PACK-POB, SKU DYNJ33520A; 4) SINUS PACK, SKU D…
- Z-0721-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A
- Z-0722-2025 · Open, ClassifiedSEPTOPLASTY PACK Convenient kit, SKU DYNJ44759F. Convenience kits used for various procedures.
- Z-0723-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: 1) CRANIOTOMY, SKU CDS982719X; 2) MVFF/FIBU #54-RF, SKU DYNJ21927Q; 3) MVFF/FIBU #54-RF, SKU DYNJ21927S; 3) NEURO PED…
- Z-0724-2025 · Open, ClassifiedMedline brand, medical procedure convenience kits, labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119T; 2) NEURO KIT, SKU CDS981327Q; 3) NEURO POSTERIOR SPINE PACK-LF, SKU…
- Z-0725-2025 · Open, ClassifiedKIT,SPINE LH Convenience kit, SKU DYNJ906448A.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page