Access hsTnI Reagent, REF B52699, Immunoassay Method… recallZ-0727-2025 · Beckman Coulter, Inc. · initiated Nov 18, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.

FDA root cause category: Component design/selection

What the company is doing

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 11/18/2024 via USPS and email. The notice explained the issue, impact to patient care, and requested the following actions be taken: " - Follow your established laboratory protocols for analyzing and retesting discrepant samples if an observed Access hsTnI test result does not align with the patient s clinical presentation. - Beckman Coulter recommends reviewing the content of this letter with your laboratory and/or medical director to determine appropriate next steps. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded the affected product listed above to another laboratory, please provide them a copy of this letter."

Recall details

Recall number
Z-0727-2025
Status
Open, Classified
Initiated by the company
Nov 18, 2024
Posted by the FDA
Dec 20, 2024
Recalling firm
Beckman Coulter, Inc.
Quantity
2,109,804 units
Product code
MMI
FDA submissions
K121214, K172783, K222881, K230648
Distribution
Worldwide

Product description

Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.

Code information (models, lots, serial numbers)

UDI/DI 15099590693183, All Lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.